Titel Uncertainty contribution of surface form errors in refractive index measurement of solid prisms
Autor(in); Institution
Schake, Markus; 4.2, Bild- und Wellenoptik, PTB-Braunschweig
Quelle/Jahr DGaO Proceedings 2026:(2026), 1 - 2
Availability [online only]
ISSN 1614-8436 (online)
DOI
URL
Konferenzangaben 127. Jahrestagung der Deutschen Gesellschaft für angewandte Optik (DGaO), Hamburg, 26, May - 29, May, 2026, Deutschland
Freie Schlagworte refractive index ; uncertainty ; solid prism ; surface form
Zusammenfassung The Physikalisch-Techniche Bundesanstalt (PTB) in Braunschweig offers a calibration service for the refractive index of solid prisms. Such calibrated prisms are used as reference standards for refractometers. At PTB a combination of the prism’s apex angle and its minimum deviation angle is employed to determine the refractive index of the material. The measurement uncertainty of the refractive index measurement is directly linked to the uncertainty of the apex angle and the angle of minimum deviation. Therefore, surface form errors on the prism’s transmitting surfaces impact the measurement uncertainty in refractive index measurement with this method. This contribution presents a new modelling concept to estimate the refractive index uncertainty contribution of surface form errors, which are determined by a Fizeau interferometer.
Themenbereich der Metrologie Länge, dimensionelle Metrologie

Zitierung

Schake, M. (2026). Uncertainty contribution of surface form errors in refractive index measurement of solid prisms. 127. Jahrestagung der Deutschen Gesellschaft für angewandte Optik (DGaO), Hamburg, 26, May - 29, May, 2026, Deutschland. https://doi.org/10.83229/127-a004