Zugriffsnummer 56519
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel Open-source Python-Based Digital Metrological Twin of Array Spectroradiometers Enabling Their Use as Transfer Standard Instruments
Autor(in); Institution
Hinken, David; Institut für Solarenergieforschung Hameln (ISFH), Emmerthal, GERMANY
Brinkmann, Malte; Institut für Solarenergieforschung Hameln (ISFH), Emmerthal, GERMANY
Vijeta, ; 4.1, Photometrie und Spektroradiometrie, PTB-Braunschweig
Sperfeld, Peter; 4.1, Photometrie und Spektroradiometrie, PTB-Braunschweig
van Dijk, Marcel; VSL, Delft, THE NETHERLANDS
Seymour, Aaron; VSL, Delft, THE NETHERLANDS
Rammeloo, Clemens; National Physical Laboratory (NPL), Teddington, UNITED KINGDOM
Zuber, Ralf; Gigahertz Optik GmbH, Türkenfeld, GERMANY
Nevas, Saulius; 4.1, Photometrie und Spektroradiometrie, PTB-Braunschweig
Quelle/Jahr The 16th International Conference on New Developments and Applications in Optical Radiometry : Book of Abstracts:(2026), 94 - 95
URL
Konferenzangaben The 16th International Conference on New Developments and Applications in Optical Radiometry ; NEWRAD 2026, Seogwipo, Jeju Island, 13-17, September, 2026, South Korea
Zusammenfassung We present an open-source Python-based implementation of a digital metrological twin of array spectroradiometers. The software converts raw digital spectroradiometer output – pixel numbers and digital counts – into radiometric and photometric quantities with fully assigned measurement uncertainties. Starting from detector signals in pixel space, the software corrects for linearity, straylight, optical bandwidth, spectral responsivity and pixel number to wavelength transfer. The built-in Monte Carlo approach calculates measurement uncertainties, including correlations in terms of wavelength and other device properties. As a result, spectral and spectrally integrated quantities are obtained in physically meaningful units with full measurement uncertainty, enabling array spectroradiometers to be used as transfer standard instruments in spectroradiometry. We validate the digital metrological twin with synthetic data sets from a simulation of a spectroradiometer and demonstrate its usage with practical examples. In addition, we discuss the importance of correlations when calculating measurement uncertainties of spectrally integrated values.
Forschungsprojekt 22IEM05: NEWSTAND: New Calibration Standards and Methods for Radiometry and Photometry After Phaseout of Incandescent Lamps
Förderinformationen (1) Förderername: EURAMET
Förderprogramm: EPM 2022
Titel der Förderung: 22IEM05: NEWSTAND: New Calibration Standards and Methods for Radiometry and Photometry After Phaseout of Incandescent Lamps
URI der Förderung: https://www.ptb.de/epm2022/newstand/home

Zitierung

Hinken, D., Brinkmann, M., Vijeta, Sperfeld, P., van Dijk, M., Seymour, A., Rammeloo, C., Zuber, R., & Nevas, S. (2026). Open-source Python-Based Digital Metrological Twin of Array Spectroradiometers Enabling Their Use as Transfer Standard Instruments. The 16th International Conference on New Developments and Applications in Optical Radiometry ; NEWRAD 2026, Seogwipo, Jeju Island, 13-17, September, 2026, South Korea.