Titel Towards standardizing low-field MRI: Multi-parametric phantom comparison across sites
Autor(in); Institution
Herthum, Helge; Berlin Center for Advanced Neuroimaging, Charité – Universitätsmedizin Berlin, GERMANY
van den Broek, Ruben; Department of Radiology, C.J. Gorter MRI Center, Leiden University Medical Center, THE NETHERLANDS
Schote, David; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin
Kulikov, Ilia; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin
Frintz, Jan Gregor; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin
Hetzer, Stefan; Berlin Center for Advanced Neuroimaging, Charité – Universitätsmedizin Berlin, GERMANY
Redshaw Kranich, Catarina; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin
Webb, Andrew; C.J. Gorter MRI Center, Leiden University Medical Center, THE NETHERLANDS
Kolbitsch, Christoph; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin
Winter, Lukas; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin; Open Source Imaging Initiative (OSI²) e.V., Berlin, GERMANY
Quelle/Jahr Proceedings of the International Society for Magnetic Resonance in Medicine:(2026)
Availability [online only]
ISSN 1065-9889 (online)
DOI
Verlag ISMRM
Konferenzangaben Annual Meeting of the ISMRM, Capetown, 9-14, Mai, 2026, South Africa
Freie Schlagworte Phantoms ; Pulse ; Sequence ; Design ; Quantitative ; MRI Ultra-Low-Field ; MRI Image Quality Assessment
Zusammenfassung Motivation: Low-field MRI is promising for accessible imaging, but variability across systems poses a major challenge. Goals: To assess inter-site consistency in image quality and quantitative metrics using the same phantom. Approach: Identical phantoms were scanned at two sites using similar 46 mT MRI systems. Structural and T1/T2 mapping data were compared for SNR, uniformity, geometric accuracy, and relaxation times. Results: Despite hardware differences and image quality variations, quantitative mapping showed very good agreement. However, site-specific artifacts highlighted the need for system characterization. Phantom-based evaluations can guide harmonization efforts and improve reproducibility in low-field MRI across platforms.

Zitierung

Herthum, H., van den Broek, R., Schote, D., Kulikov, I., Frintz, J. G., Hetzer, S., Redshaw Kranich, C., Webb, A., Kolbitsch, C., & Winter, L. (2026). Towards standardizing low-field MRI: Multi-parametric phantom comparison across sites. Annual Meeting of the ISMRM, Capetown, 9-14, Mai, 2026, South Africa. https://archive.ismrm.org/