Datensatz noch nicht abgeschlossen
Zugriffsnummer 56034
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel Evaluation of implant-induced SAR in low-field MRI (B0 = 50 mT) using representative geometries
Autor(in); Institution
Ernst, Ariane; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin
Winter, Lukas; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin; Open Source Imaging Initiative (OSI²) e.V., Berlin, GERMANY
Zanovello, Umberto; Istituto Nazionale di Ricerca Metrologica, Turin, ITALY
Seifert, Frank; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin
Quelle/Jahr Proceedings of the International Society for Magnetic Resonance in Medicine:(2026)
Availability [online only]
ISSN 1065-9889 (online)
URL
Verlag ISMRM
Konferenzangaben Annual Meeting of the ISMRM, Capetown, 9-14, Mai, 2026, South Africa
Zusammenfassung Recent advances in magnet design, image reconstruction, and deep learning have renewed interest in low-field (LF) MRI for portable neuroimaging. Modern systems achieve faster acquisitions and improved image quality, showing diagnostic promise for neurological disorders and safety benefits such as reduced magnetic forces and lower SAR. However, simulations and phantom studies have shown that localized heating in implants at 0.55 T can exceed thatat 1.5 T in particular scenarios, cautioning against blanket safety assumptions across LF systems .

Zitierung

Ernst, A., Winter, L., Zanovello, U., & Seifert, F. (2026). Evaluation of implant-induced SAR in low-field MRI (B0 = 50 mT) using representative geometries. Annual Meeting of the ISMRM, Capetown, 9-14, Mai, 2026, South Africa.