| Zugriffsnummer | 55895 |
| Dokumenttyp | Konferenzartikel |
| Sprache | Englisch |
| Titel | Room temperature self-calibration of photodiodes reaching standard uncertainties of 34 ppm |
| Autor(in); Institution |
Heitmann Solheim, Johanne; Justervesenet JV, Kjeller, NORWAY
Bardalen, Eivind; University of South-Eastern Norway, Borre, NORWAY
Pohl, Tobias; 7.3, Detektorradiometrie und Strahlungsthermometrie, PTB-Berlin
Paunov, Masa; Justervesenet JV, Kjeller, NORWAY
Gran, Jarle; Justervesenet JV, Kjeller, NORWAY
|
| Quelle/Jahr | The 16th International Conference on New Developments and Applications in Optical Radiometry : Book of Abstracts:(2026), 49 - 50 |
| URL | |
| Konferenzangaben | NEWRAD 2026 - 16th International Conference on New Developments and Applications in Optical Radiometry, Seogwipo, 13-17, September, 2026, South Korea |
| Zusammenfassung | A new dual‑mode module design was developed and tested, based on COMSOL simulations to ensure sufficient thermal equivalence between the two heating modes. The modules exhibit low sensitivity to beam alignment, confirming the robustness of the design. The measurements revealed that the apparent internal quantum deficiency is dependent on optical power, something which was not predicted by the simulations. However, applying a correction for this dependence yields good agreement, with a record‑low measurement uncertainty of 34 ppm absolute. |
| Themenbereich der Metrologie | Photometrie und Radiometrie |
| Innovationscluster | Umwelt und Klima |
| Geschäftsfelder | Grundlagen der Metrologie |
| Förderinformationen (1) |
Förderername: European Partnership on Metrology
Titel der Förderung: Scale-up Förderungsnummer: 22IEM06 |
Zitierung
Heitmann Solheim, J., Bardalen, E., Pohl, T., Paunov, M., & Gran, J. (2026). Room temperature self-calibration of photodiodes reaching standard uncertainties of 34 ppm. NEWRAD 2026 - 16th International Conference on New Developments and Applications in Optical Radiometry, Seogwipo, 13-17, September, 2026, South Korea.