Titel Round robin comparison of power losses performed by Epstein frame and sst above 50 hz at room temperature
Autor(in); Institution
Ulvr, Michal; Czech Metrology Institute (CMI), Brno, CZECH REPUBLIC
Weickert, Franziska; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Pfnür, Korbinian; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Lüdke, Joachim; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Hoffmann, Katja; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Harmon, Stuart; National Physical Laboratory (NPL), Teddington, UK
Brunt, Daniel; National Physical Laboratory (NPL), Teddington, UK
Wilson, Adam; National Physical Laboratory (NPL), Teddington, UK
Pasquale, Massimo; Istituto Nazionale di Ricerca Metrologica (INRIM), Torino, ITALY
Appino, Carlo; Istituto Nazionale di Ricerca Metrologica (INRIM), Torino, ITALY
Gerada, Chris; University of Nottingham (UNOTT), Nottingham, UK
Ramanathan Mathavan Jeyabalan, Ram; University of Nottingham (UNOTT), Nottingham, UK
Vakil, Gaurang; University of Nottingham (UNOTT), Nottingham, UK
Quelle/Jahr Journal of Magnetism and Magnetic Materials: 642 (2026), 1 - 7
Artikelnummer 173897
ISSN 0304-8853 ; 1873-4766
DOI
Verlag Amsterdam: Elsevier
Freie Schlagworte Epstein ; power loss ; reference sample ; SST ; comparison
Zusammenfassung Epstein frame and single sheet tester are two methods used for measurement of power losses of the electrical steel sheets or strips. Both methods and the setups are described in the IEC standards. Each national metrology institute or other metrological laboratory has different setups for power losses measurement and the validation of its accuracy and parameters can only be done by comparisons. So far, every performed comparison was carried out only at 50 Hz or 60 Hz. Within the EMPIR project "HEFMAG", improved metrological infrastructure for the determination of power losses using Epstein frame at induction values close to saturation and at frequency ranges between 2 kHz and to 10 kHz was built. To validate the improved setups, a round robin comparison of power loss measurements was conducted by five laboratories. The results of the round robin comparison are discussed including measurement uncertainties.
Kostenfreier Zugang Open Access Hybrid
Rechteinformation CC BY-NC-ND 4.0 ; Creative Commons Attribution NonCommercial NoDerivatives 4.0 License
Themenbereich der Metrologie Elektrizität und Magnetismus
Forschungsprojekt 19ENG06: HEFMAG: Metrology of magnetic losses in electrical steel sheets for high-efficiency energy conversion
Förderinformationen (1) Förderername: European Commission (EC)
Förderer ID: 0000 0001 2242 8989
Förderer ID Typ: ISNI
Förderprogramm: EMPIR 2019 Energy
Titel der Förderung: 19ENG06: HEFMAG: Metrology of magnetic losses in electrical steel sheets for high-efficiency energy conversion
Förderungsnummer: 19ENG06

Zitierung

Ulvr, M., Weickert, F., Pfnür, K., Lüdke, J., Hoffmann, K., Harmon, S., Brunt, D., Wilson, A., Pasquale, M., Appino, C., Gerada, C., Ramanathan Mathavan Jeyabalan, R., & Vakil, G. (2026). Round robin comparison of power losses performed by Epstein frame and sst above 50 hz at room temperature. Journal of Magnetism and Magnetic Materials, 642, 1–7. https://doi.org/10.1016/j.jmmm.2026.173897