Titel Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the PTB, Germany, and the BIPM in mammography x-rays
Autor(in); Institution
Kessler, C.; Bureau International des Poids et Mesures, Sèvres Cedex, FRANCE
Burns, D.; Bureau International des Poids et Mesures, Sèvres Cedex, FRANCE
Pojtinger, Stefan; 6.2, Dosimetrie für Strahlentherapie und Röntgendiagnostik, PTB-Braunschweig
Quelle/Jahr Metrologia: 61 (2024), 1A, Tech. Suppl., 13 S.
Artikelnummer 06019
Availability [online only]
ISSN 0026-1394 (print) ; 1681-7575 (online)
DOI
Verlag Bristol: IOP Publishing
Freie Schlagworte PTB high performance compute cluster ; Key comparison ; air-kerma ; free-air chamber ; primary standard
Zusammenfassung A new key comparison has been made between the air-kerma standards of the PTB, Germany, and the BIPM in mammography x-ray beams. The results show the standards to agree at the level of the expanded uncertainty of the comparison of 7.2 parts in 103. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database.