Zugriffsnummer 53599
Dokumenttyp Zeitschriftenartikel Open Access Hybrid
Peer Review mit Peer Review
Sprache Englisch
Titel Systematic study on the influence of inductive chargers on active personal dosemeters
Autor(in); Institution
Ketelhut, Steffen; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Zutz, Hayo; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Hupe, Oliver; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Quelle/Jahr Journal of Radiological Protection: 44 (2024), 4, 1 - 6
Artikelnummer 041510
ISSN 0952-4746 (print) ; 1361-6498 (online)
DOI
Verlag Bristol: IOP Publishing
Freie Schlagworte personal dosimetry ; inductive charging ; type testing ; electromagnetic interference
Zusammenfassung Electromagnetic compatibility testing plays an important role in the type testing of radiation protection dosemeters in view of technical developments and the associated increase in electromagnetic fields. Lately, the use of inductive charging devices has grown as a user-friendly type of charging mobile-phones. In this article, we investigate their impact on active personal dosemeters (APD). The measurements show a substantial additional dose reading of up to several tens of mSv when exposed to the field of an inductive charger for 20 s. According to the widely used Qi standard, the charging devices operate at frequencies of between 87 kHz and 205 kHz for power transfers between 5 W and 30 W. These parameters fall outside the scope of type-testing standards for APD. An update of the standards might therefore be necessary.
Kostenfreier Zugang Open Access Hybrid
Rechteinformation CC BY 4.0 ; Creative Commons Attribution 4.0 License

Zitierung

Ketelhut, S., Zutz, H., & Hupe, O. (2024). Systematic study on the influence of inductive chargers on active personal dosemeters. Journal of Radiological Protection, 44(4), 1–6. https://doi.org/10.1088/1361-6498/ad8fb7

Exportieren