Titel Hyperpolarized 129Xe for benchtop NMR: Inline instrumentation for automated yet flexible operation yielding high polarization
Autor(in); Institution
Kilian, Wolfgang; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin
Gulich, Samira; 8.3, Biomedizinische Optik, PTB-Berlin
Riemer, Thomas; 8.3, Biomedizinische Optik, PTB-Berlin; University of Leipzig, Medical Faculty, Institute of Medical Physics and Biophysics, Leipzig, GERMANY
Mitschang, Lorenz; 8.3, Biomedizinische Optik, PTB-Berlin
Quelle/Jahr Journal of Magnetic Resonance Open: 22 (2025)
Artikelnummer 100179
ISSN 2666-4410
DOI
Verlag New York, NY [u.a.]: Elsevier
Freie Schlagworte Benchtop NMR ; Hyperpolarized 129Xe ; Absolute polarization ; In-situ sample gas saturation
Zusammenfassung The use of hyperpolarized media in combination with benchtop NMR spectrometers is currently extensively investigated to enable cost effective but highly sensitive applications. In this work, the instrumentation for the use of hyperpolarized 129Xe gas on a commercial benchtop-NMR spectrometer in a fully controllable yet automated mode is presented. A continuous-flow 129Xe polarizer is operated inline with the benchtop spectrometer. While the adjustment of the xenon gas partial pressure in the range of mbar to bar as well as the volume flow through the NMR sample is facilitated by the polarizer’s mass-flow controllers and two back-pressure regulators, respectively, the gas flow towards the NMR sample in situ in the benchtop magnet is time-controlled within the spectrometer’s RF pulse sequence programming. A calibration procedure for the gas flow control as well as thermally polarized xenon gas standards are introduced for quantification of the absolute 129Xe polarization. In this way, the 129Xe polarization achieved in the NMR measurement is determined in the 50% to 5% regime for very lean 1 mbar to high 500 mbar xenon partial pressure, respectively – in the optimum a more than 250,000-fold increase in comparison to thermal polarization. Such instrumentation may be implemented by combining any of the well-established continuous-flow 129Xe polarizer with any of the commercial benchtop spectrometers, thus facilitating high-performance hyperpolarized 129Xe benchtop NMR studies.