Titel Comparison of S-parameter measurement methods for attenuators
Autor(in); Institution
Schramm, Andreas; 2.2, Hochfrequenz und Felder, PTB-Braunschweig
Gellersen, Frauke; 2.2, Hochfrequenz und Felder, PTB-Braunschweig
Kuhlmann, Karsten; 2.2, Hochfrequenz und Felder, PTB-Braunschweig
Quelle/Jahr ARFTG 2024:(2024), 4 S.
Verlag Piscataway, NJ: IEEE
Konferenzangaben 103rd ARFTG Microwave Measurement Conference, Washington, DC, 21, June, 2024, USA
Freie Schlagworte calibration ; attenuation ; measurement uncertainty
Zusammenfassung In this work two methods for characterization of attenuators - meaning two-ports with high return loss value and well-defined insertion loss - in the range of 0 dB to 30 dB attenuation are simulated and compared. The main focus is the evaluation of the achievable measurement uncertainty as well as the required effort and scope of the measurement. The comparison is carried out using a coaxial attenuator type 2.92 mm in the frequency range up to 44 GHz. The first method is based on two 1-port calibrations whereas the second methods is one full 2-port calibration using the attenuator as the unknown thru.
Themenbereich der Metrologie Elektrizität und Magnetismus

Zitierung

Schramm, A., Gellersen, F., & Kuhlmann, K. (2024). Comparison of S-parameter measurement methods for attenuators. 103rd ARFTG Microwave Measurement Conference, Washington, DC, 21, June, 2024, USA.