Cardiac function and strain analysis at 7T in comparison to 3T
Autor(in); Institution
Jung, Bernd; Dept. of Diagnostic, Interventional and Pediatric Radiology, Inselspital, Bern University Hospital, Bern, SWITZERLAND; Translation Imaging Center (TIC), Swiss Institute for Translational and Entrepreneurial Medicine, Bern, SWITZERLAND
Binanno, Gabrielle; Swiss Institute for Translational and Entrepreneurial Medicine, Bern, SWITZERLAND; Advanced Clinical Imaging Technology, Siemens Healthineers International AG, Bern, SWITZERLAND
Schmitter, Sebastian; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin; Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, UNITED STATES
Guensch, Dominik P; Dept. of Diagnostic, Interventional and Pediatric Radiology, Inselspital, Bern University Hospital, Bern, SWITZERLAND; Department of Anaesthesiology and Pain Medicine, Inselspital, Bern University Hospital, Bern, SWITZERLAND
Fischer, Kady; Department of Anaesthesiology and Pain Medicine, Inselspital, Bern University Hospital, Bern, SWITZERLAND
Quelle/Jahr
Proceedings of the International Society for Magnetic Resonance in Medicine:(2024), 1
- 2
Myocardium ; High-Field MRI ; Cardiac function ; Strain
Zusammenfassung
Motivation: To investigate whether a single set of shim values can provide reliable image quality for cardiac MRI at 7T.
Goals: To establish an e!cient workflow for cardiac cine imaging at 7T.
Approach: Short-axis and long-axis views were acquired in five subjects with di"erent spatial resolution at 3T and 7T. Volumetric parameters and peak global longitudinal and circumferential strain were assessed for left and right ventricle.
Results: No significant di"erences of volume parameters, field strength or spatial resolution were found. A single B1 shim set for the whole examination providing reliable image quality and quantification results constitute an essential step towards clinical cardiac imaging.
Impact: An e!cient workflow and reliable image quality of cine images for the assessment of cardiac volumes and strain at 7 Tesla could be realized with results in close agreement to 3T – an essential step towards clinical routine cardiac examinations.
Zitierung
Jung, B., Binanno, G., Schmitter, S., Guensch, D. P., & Fischer, K. (2024). Cardiac function and strain analysis at 7T in comparison to 3T. ISMRM & ISMRT Annual Meeting & Exhibition, Singapore, 04-09, Mai, 2024, Singapore. https://doi.org/10.58530/2024/1657