Zugriffsnummer 51904
Dokumenttyp Zeitschriftenartikel Open Access Gold
Peer Review mit Peer Review
Sprache Englisch
Titel Self-calibration strategies for reducing systematic slope measurement errors of autocollimators in deflectometric profilometry
Autor(in); Institution
Geckeler, Ralf D.; 5.2, Dimensionelle Nanometrologie, PTB-Braunschweig
Just, Andreas; 5.2, Dimensionelle Nanometrologie, PTB-Braunschweig
Krause, Michael; 5.2, Dimensionelle Nanometrologie, PTB-Braunschweig
Schnabel, Olaf; Schnabel Elektronische Messtechnik, Wedel, GERMANY
Lacey, Ian; Advanced Light Source, Lawrence Berkeley National Laboratory, Berkeley, CA, USA
English, Damon; Engineering Division, Lawrence Berkeley National Laboratory, Berkeley, CA, USA
Yashchuk, Valeriy V.; Advanced Light Source, Lawrence Berkeley National Laboratory, Berkeley, CA, USA
Quelle/Jahr Journal of Synchrotron Radiation: 31 (2024) , 4, 1 - 11
ISSN 0909-0495 (print) ; 1600-5775 (online)
DOI
Verlag Oxford: Blackwell
Freie Schlagworte angle metrology ; optical metrology ; autocollimator ; form measurement ; deflectometry ; profilometry ; data processing ; error suppression ; synchrotron ; free-electron laser
Zusammenfassung Deflectometric profilometers are used to precisely measure the form of beam shaping optics of synchrotrons and X-ray free-electron lasers. They often utilise autocollimators which measure slope by evaluating the displacement of a reticle image on a detector. Based on our privileged access to the raw image data of an autocollimator, we discuss novel strategies to reduce the systematic measurement errors by using a set of overlapping images of the reticle obtained at different positions on the detector. We demonstrate that imaging properties such as, for example, geometrical distortions and vignetting, can be extracted from this redundant set of images without recourse to external calibration facilities. This approach is based on the fact that the properties of the reticle itself do not change, all changes in the reticle image are due to the imaging process. Firstly, by combining interpolation and correlation, we are able to determine the shift of a reticle image relative to a reference image with minimal error propagation. Secondly, the intensity of the reticle image is analysed as a function of its position on the CCD and a vignetting correction is calculated. Thirdly, the size of the reticle image is analysed as a function of its position and an imaging distortion correction is derived. We demonstrate that, for different measurement ranges and aperture diameters of the autocollimator, reductions in the systematic errors of up to a factor of 4-5 can be achieved without recourse to external measurements.
Kostenfreier Zugang Open Access Gold
Rechteinformation CC BY 4.0 ; Creative Commons Attribution 4.0 License

Zitierung

Geckeler, R. D., Just, A., Krause, M., Schnabel, O., Lacey, I., English, D., & Yashchuk, V. V. (2024). Self-calibration strategies for reducing systematic slope measurement errors of autocollimators in deflectometric profilometry. Journal of Synchrotron Radiation, 31(4), 1–11. https://doi.org/10.1107/S1600577524003552