Zugriffsnummer 51229
Dokumenttyp Zeitschriftenartikel Open Access Hybrid
Peer Review mit Peer Review
Sprache Englisch
Titel Realization of the absorbed dose to water for 70kV electronic brachytherapy sources for surface applications
Autor(in); Institution
Schneider, Thorsten; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Quelle/Jahr Physics in Medicine and Biology: 68 (2023), 24, 1 - 11
Artikelnummer 245022
ISSN 0031-9155 (print) ; 1361-6560 (online)
DOI
Verlag Bristol: IOP Publishing
Freie Schlagworte electronic brachytherapy ; dosimetry ; surface treatment
Zusammenfassung Objective. The purpose of this work is to establish a primary standard for the absorbed dose to water for surface applications with 70 kV electronic brachytherapy sources and thereby to investigate several reference conditions with respect to applicability in metrology institutes, calibration labs and clinics. Approach. A primary standard for the absorbed dose to water for LDR-Seeds (ipFAC) was utilized. For this, the method of evaluation was modified to account for the different geometries in interstitial and surface brachytherapy and for the different energy distributions of the radiation fields, respectively. The correction factors required to determine the absorbed dose to water were evaluated using Monte Carlo (MC) simulations. MC-calculations were also used to estimate the uncertainties of this method. Main results. It could be shown that determining the absorbed dose to water in 1 cm depth below the surface of a water phantom Dw (1 cm) is feasible with the ipFAC. Thereby the method to determine Dw (1 cm) when the source is placed at 50 cm distance and the beam is collimated turns out to be more robust to variations in the directional emittance of the source than placing the source applicator assembly directly on the phantom. Significance. The first method features significantly lower uncertainties, 1.4% compared to 3.7% (both for k = 2), for the second one. However, a calibration based on a transfer chamber calibrated with the second method is more practical and easier to implement in clinical routine.
Kostenfreier Zugang Open Access Hybrid
Rechteinformation CC BY 4.0 ; Creative Commons Attribution 4.0 License
Themenbereich der Metrologie Ionisierende Strahlung
Innovationscluster Gesundheit
Geschäftsfelder Grundlagen der Metrologie
Forschungsprojekt 18NRM02: PRISM-eBT: Primary standards and traceable measurement methods for X-ray emitting electronic brachytherapy devices
Förderinformationen (1) Förderername: European Commission (EC)
Förderer ID: 0000 0001 2242 8989
Förderer ID Typ: ISNI
Förderprogramm: EMPIR 2018 Normative
Titel der Förderung: 18NRM02: PRISM-eBT: Primary standards and traceable measurement methods for X-ray emitting electronic brachytherapy devices
Förderungsnummer: 18NRM02

Zitierung

Schneider, T. (2023). Realization of the absorbed dose to water for 70kV electronic brachytherapy sources for surface applications. Physics in Medicine and Biology, 68(24), 1–11. https://doi.org/10.1088/1361-6560/ad0d41

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag