| Titel | Bilateral comparison between viscosity measurement standard systems developed by KRISS and PTB |
| Autor(in); Institution |
Lee, Seok Hwan; Division of Physical Metrology, Korea Research Institute of Standards and Science,Yuseong-Gu, Daejeon, REPUBLIC OF KOREA
Rauch, Jürgen; 3.3, Physikalische Chemie, PTB-Braunschweig
Nam, Hyun-Seung; Division of Physical Metrology, Korea Research Institute of Standards and Science,Yuseong-Gu, Daejeon, REPUBLIC OF KOREA
Yoon, Byung-Ro; Division of Physical Metrology, Korea Research Institute of Standards and Science,Yuseong-Gu, Daejeon, REPUBLIC OF KOREA
Kang, Woong; Division of Physical Metrology, Korea Research Institute of Standards and Science,Yuseong-Gu, Daejeon, REPUBLIC OF KOREA
Kwon, Suyong; Division of Physical Metrology, Korea Research Institute of Standards and Science,Yuseong-Gu, Daejeon, REPUBLIC OF KOREA
|
| Quelle/Jahr | Measurement: 218 (2023), 1 - 7 |
| Artikelnummer | 113143 |
| ISSN | 0263-2241 (print) ; 1873-412X (online) |
| DOI | |
| Verlag | Amsterdam: Elsevier |
| Freie Schlagworte | viscosity measurement standard ; bilateral comparison ; capillary viscometer ; uncertainty evaluation |
| Zusammenfassung | The viscosity measurement standard systems developed by the Korea Research Institute of Standards and Science (KRISS) and Physikalisch-Technische Bundesanstalt (PTB) were compared, and the international equivalence of the standard viscosity measurement system built by KRISS was confirmed. In the system developed by KRISS, the measurement range of the viscosity measurement standard system was 0.3-100000 mm2/s with 0.17-0.5% uncertainty (U, k = 2). The density of a liquid was measured using a pycnometer, and an absolute viscosity of 0.3-8000 mPas was measured with an uncertainty of 0.18-0.6%. The viscosity measurement standard systems of KRISS and PTB were bilaterally compared, using three different viscosity standard liquids synthesized by PTB, under six experimental conditions. Considering the uncertainty, the calculated En was < 1 (0.17-0.72) for all experimental cases. Therefore, we confirmed the viscosity standard system created recently by KRISS to be significantly equivalent to that developed by PTB. |
| Kostenfreier Zugang | Open Access Hybrid |
| Rechteinformation | CC BY-NC-ND 4.0 ; Creative Commons Attribution NonCommercial NoDerivatives 4.0 License |
| Themenbereich der Metrologie | Metrologie in der Chemie und Stoffeigenschaften |
| Förderinformationen (1) |
Förderername: Government of the Republic of Korea Ministry of Trade Industry and Energy (MOTIE)
Förderer ID: 0000 0004 0381 8086 Förderer ID Typ: ISNI Förderungsnummer: 20212020800070 |
| Förderinformationen (2) |
Förderername: Korea Ministry of Science and ICT (MSIT)
Förderer ID: 0000 0004 7781 685X Förderer ID Typ: ISNI Förderungsnummer: 23011055 |