Kleinheubacher Tagung 2022, Miltenberg, 27-29, September, 2022, Deutschland
Freie Schlagworte
direct comparison transfer ; generalized efficiency ; millimeter wave power measurement ; waveguide thermoelectric sensor
Zusammenfassung
This paper describes the determination and application of the generalized efficiency of waveguide thermoelectric power sensors in the direct comparison transfer method for the frequency range of 110 GHz to 170 GHz. Since waveguides cannot support DC signals, the generalized efficiency is required for waveguide thermoelectric sensors. The thermoelectric sensor with the generalized efficiency can be used as transfer standard to calibrate all types of power sensor in the direct comparison transfer, e.g., to establish the measurement traceability of commercial power meters for millimeter wave frequencies.
Themenbereich der Metrologie
Elektrizität und Magnetismus
Zitierung
Perangin-Angin, W. K., Kuhlmann, K., Rühaak, J., & Phung, G. N. (2022). Generalized efficiency of waveguide thermoelectric power sensor. Kleinheubacher Tagung 2022, Miltenberg, 27-29, September, 2022, Deutschland.