Zugriffsnummer 48646
Dokumenttyp Konferenzartikel
Peer Review unbekannt
Sprache Englisch
Titel Static and dynamic parallel transmission (pTx) for human cardiac MRI at 14.0 T
Autor(in); Institution
Nurzed, B.; Berlin Ultrahigh Field Facility, MDC, Berlin, GERMANY
Eigentler, T.; Berlin Ultrahigh Field Facility, MDC, Berlin, GERMANY
Aigner, Christoph; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin
Schmitter, Sebastian; 8.1, Biomedizinische Magnetresonanz, PTB-Berlin
Niendorf, T.; Berlin Ultrahigh Field Facility, MDC, Berlin, GERMANY
Quelle/Jahr Proceedings of the International Society for Magnetic Resonance in Medicine: 30 (2022), 2 S.
Artikelnummer 1120
Availability [online only]
ISSN 1545-4428 (ONLINE)
DOI
Konferenzangaben 31st Joint Annual Meeting ISMRM-ESMRMB & ISMRT, London, 07-12, May, 2022, UK
Zusammenfassung Transmission field inhomogeneities at ultrahigh and extreme field MRI can be offset by using static or dynamic pTx. Responding to the challenges and recognizing the opportunities of cardiac MRI, this abstract examines the feasibility of parallel transmission (pTx) using fractionated dipole (FRD) RF array configurations for static and dynamic B1+ homogenization of the heart at 7.0T and 14.0T. Our results reveal that static pTx provides limited performance at 14.0 T but dynamic pTx enables uniform excitation of the heart at 14.0T. This finding is heartening and provides the technical foundation for explorations into cardiac MRI at 14.0T.
Themenbereich der Metrologie Metrologie in der Medizin

Zitierung

Nurzed, B., Eigentler, T., Aigner, C., Schmitter, S., & Niendorf, T. (2022). Static and dynamic parallel transmission (pTx) for human cardiac MRI at 14.0 T. 31st Joint Annual Meeting ISMRM-ESMRMB & ISMRT, London, 07-12, May, 2022, UK. https://doi.org/10.58530/2022/1120