Zugriffsnummer 47807
Dokumenttyp Zeitschriftenartikel Open Access Hybrid
Peer Review mit Peer Review
Sprache Englisch
Titel Radiation measurements and simulations at the PHELIX laser facility, GSI
Autor(in); Institution
Behrens, Rolf; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Hufschmidt, P.; Erlangen Centre for Astroparticle Physics, Erlangen, GERMANY
Roth, Jürgen; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Akstaller, B.; Erlangen Centre for Astroparticle Physics, Erlangen, GERMANY
Haag, D.; Erlangen Centre for Astroparticle Physics, Erlangen, GERMANY
Hupe, Oliver; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Schmidt, S.; Erlangen Centre for Astroparticle Physics, Erlangen, GERMANY
Schreiner, S.; Erlangen Centre for Astroparticle Physics, Erlangen, GERMANY
Michel, T.; Erlangen Centre for Astroparticle Physics, Erlangen, GERMANY
Quelle/Jahr Journal of Instrumentation: 17 (2022)
Artikelnummer P05033
Availability [online only]
ISSN 1748-0221 (ONLINE)
DOI
Verlag London: IOP Publishing
Freie Schlagworte Analysis and statistical methods ; Detector modelling and simulations I (interaction of radiation with matter, interaction of photons with matter, interaction of hadrons with matter, etc) ; Dosimetry concepts and apparatus ; Nuclear instruments and methods for hot plasma diagnostics
Zusammenfassung The absolute spectral electron and photon emission from a Tungsten target irradiated by the Petawatt High-Energy Laser for Heavy Ion EXperiments (PHELIX) facility was measured inside the PHELIX target chamber. For this, a thermo-lumi-nescence detector (TLD)-based few-channel spectrometer (FCS) was used in combination with a Bayesian data analysis. The results of the FCS measurement served as input for a Monte-Carlo particle transport simulation to determine the spectra and dose out of the PHELIX target chamber. Two different simulation methods proved to be consistent and, therefore, reliable. The simulation results were compared to the dose measured with a dosemeter that consists of three Dosepix detectors in combination with a neural network analysis. The two dose results show a satisfactory agreement.
Kostenfreier Zugang Open Access Hybrid
Rechteinformation CC BY 4.0 ; Creative Commons Attribution 4.0 License
Themenbereich der Metrologie Ionisierende Strahlung