Zugriffsnummer 47751
Dokumenttyp Zeitschriftenartikel Open Access Hybrid
Peer Review mit Peer Review
Sprache Englisch
Titel Comparison in dynamic primary calibration of digital-output accelerometer between CEM and PTB
Autor(in); Institution
Seeger, Benedikt; 1.7, Akustik und Dynamik, PTB-Braunschweig
Jugo, Ehlimana; Institute of Metrology of Bosnia and Herzegovina (IMBiH), Sarajevo, BOSNIA AND HERZEGOVINA
Bosnjakovic, A.; Institute of Metrology of Bosnia and Herzegovina (IMBiH), Sarajevo, BOSNIA AND HERZEGOVINA
Ruiz, S.; Centro Español de Metrología (CEM), Madrid, SPAIN
Bruns, Thomas; 1.7, Akustik und Dynamik, PTB-Braunschweig
Quelle/Jahr Metrologia: 59 (2022), 3, 1 - 12
Artikelnummer 035002
Availability [online only]
ISSN 1681-7575 (ONLINE)
DOI
Verlag Bristol: IOP Publishing
Zusammenfassung In the framework of the EU-funded project Met4FoF Metrology for the Factory of the Future a method was developed to perform primary magnitude and phase calibration of accelerometers. The validation of the developed associated hard- and software, which is designed to complement conventional primary calibration systems, is demonstrated based on a comparison measurement between the project partners CEM and PTB. The results show that the statistical distribution of results is only a minor contribution to uncertainty. An evaluation of the comparison based on the conventional procedure and established CMCs proofed a good consistency between the two laboratories, although the utilized sensor was (only) industrial grade quality.
Kostenfreier Zugang Open Access Hybrid
Rechteinformation CC BY 4.0 ; Creative Commons Attribution 4.0 License
Themenbereich der Metrologie Akustik, Ultraschall, Beschleunigung
Forschungsprojekt 17IND12: Met4FoF: Metrology for the Factory of the Future
Förderinformationen (1) Förderername: European Commission (EC)
Förderer ID: 0000 0001 2242 8989
Förderer ID Typ: ISNI
Förderprogramm: EMPIR 2017 Industry
Titel der Förderung: 17IND12: Met4FoF: Metrology for the Factory of the Future
Förderungsnummer: 17IND12
URI der Förderung: https://www.ptb.de/empir2018/met4fof/home/

Zitierung

Seeger, B., Jugo, E., Bosnjakovic, A., Ruiz, S., & Bruns, T. (2022). Comparison in dynamic primary calibration of digital-output accelerometer between CEM and PTB. Metrologia, 59(3), 1–12. https://doi.org/10.1088/1681-7575/ac5a5d