Zusammenfassung Using examples of the calibration of UVA radiometers, the influence of spectral mismatch will be discussed. As the spectral responsivity of so-called UVA radiometers is never spectrally flat, it will be shown that this can lead to significant differences between measurement results even of narrow-band UVA sources typically used for calibrations. This should be taken into account both for routine calibrations of UVA radiometers and for carrying out intercomparisons.