Zugriffsnummer 45991
Dokumenttyp Zeitschriftenartikel Open Access Gold
Peer Review mit Peer Review
Sprache Englisch
Titel Development of decahydronaphthalene reference material for low flash point measurements
Autor(in); Institution
Ahmed, Hany Elmenyawi; National Institute of Standards, Giza, EGYPT
Tahoun, Ibrahim F.; National Institute of Standards, Giza, EGYPT
Zakel, Sabine; 3.7, Grundlagen des Explosionsschutzes, PTB-Braunschweig
Quelle/Jahr Egyptian Journal of Petroleum: 30 (2021), 1, 7 - 10
ISSN 1110-0621 (PRINT)
DOI
Verlag Amsterdam: Elsevier
Freie Schlagworte Flash point ; Reference Material ; Value assignment
Zusammenfassung Certified reference materials are an essential tool for calibration of testing equipment, validation of analytical methods and ensuring quality of test results. Due to lack of national low temperature reference materials, Decahydronaphthalene reference material for flash point measurements according to ISO 13736has been developed by the National Metrology Institute of Egypt (NIS). This article describes production steps of the reference material. Certification of the material was carried out by interlaboratory comparisons, where between bottles homogeneity and stability studies were conducted in explosion protection department at PTB, Germany. The interlaboratory results were statistically analyzed, and robust mean was calculated as assigned value. The certified value and expanded measurement uncertainty (k = 2) was found 55.7 °C ± 1.2 °C, respectively.
Kostenfreier Zugang Open Access Gold
Rechteinformation CC BY-NC-ND 4.0 ; Creative Commons Attribution NonCommercial NoDerivatives 4.0 License
Themenbereich der Metrologie Physikalische Sicherheitstechnik, Explosionsschutz

Zitierung

Ahmed, H. E., Tahoun, I. F., & Zakel, S. (2021). Development of decahydronaphthalene reference material for low flash point measurements. Egyptian Journal of Petroleum, 30(1), 7–10. https://doi.org/10.1016/j.ejpe.2020.12.002

Exportieren