Zugriffsnummer 45705
Dokumenttyp Zeitschriftenartikel
Peer Review mit Peer Review
Sprache Englisch
Titel Fast and robust optically pumped cesium magnetometer
Autor(in); Institution
Lebedev, Victor; 8.2, Biosignale, PTB-Berlin
Hartwig, Stefan; 8.2, Biosignale, PTB-Berlin
Middelmann, Thomas; 8.2, Biosignale, PTB-Berlin
Quelle/Jahr Advanced Optical Technologies: 9 (2020), 5, 275 - 286
ISSN 2192-8576 (PRINT) ; 2192-8584 (ONLINE)
DOI
Verlag Berlin: De Gruyter
Freie Schlagworte high speed ; optically detected magnetic resonance ; optically pumped magnetometer ; phase feedback ; unshielded operation
Zusammenfassung We present a fast and robust optically pumped magnetometer that is based on a feedback-controlled spin ensemble of cesium atoms in spin-polarized vapor. The table-top system is intended for operation in unshielded environment, and its design allows conversion into a handheld sensor head. Under strongly disturbed environmental conditions in the laboratory, the sensor exhibits a speed of more than 56 kHz, while having a slew rate of 39mT/s and a full dynamic range of 10 - 120 μT. Under these conditions a sensitivity of 33 pT/√Hz is reached. By reducing the speed to 3.6 kHz the sensitivity can be improved to 6pT/√Hz, which is close to the shot noise limit. We describe the sensor design and its optimization and demonstrate the performance of the sensor under conditions appropriate for magnetic susceptometry measurements.
Themenbereich der Metrologie Metrologie in der Medizin

Zitierung

Lebedev, V., Hartwig, S., & Middelmann, T. (2020). Fast and robust optically pumped cesium magnetometer. Advanced Optical Technologies, 9(5), 275–286. https://doi.org/10.1515/aot-2020-0024

Exportieren