Titel Laser-Doppler-Vibrometer calibration by laser stimulation
Autor(in); Institution
Volkers, Henrik; 1.7, Akustik und Dynamik, PTB-Braunschweig
Bruns, Thomas; 1.7, Akustik und Dynamik, PTB-Braunschweig
Quelle/Jahr Special issue for TC3, TC5, TC16 and TC22. Acta IMEKO: 9 (2020), 5, 357 - 360
Availability [online only]
ISSN 2221-870X
DOI
Verlag Braunschweig:
Freie Schlagworte primary calibration ; laser doppler vibrometer ; LDV calibration
Zusammenfassung A new set-up for primary laser vibrometer calibration was developed and tested at the acceleration laboratory of PTB. Contrary to existing set-ups, this configuration makes use of electro-optical excitation.While avoiding the limitations imposed by mechanical motion generators in classic set-ups, the new method still encompasses all components of commercial laser vibrometers in the calibration and thus goes beyond the current capabilities of the purely electrical excitation schemes.
Kostenfreier Zugang Open Access Gold
Rechteinformation CC BY 3.0 ; Creative Commons Attribution 3.0 License
Themenbereich der Metrologie Akustik, Ultraschall, Beschleunigung

Zitierung

Volkers, H. & Bruns, T. (2020). Laser-Doppler-Vibrometer calibration by laser stimulation. Acta IMEKO, 9(5), 357–360. http://doi.org/10.21014/acta_imeko.v9i5.1000