Zugriffsnummer 45338
Dokumenttyp Zeitschriftenartikel Open Access Gold
Peer Review mit Peer Review
Sprache Englisch
Titel Investigation of the repeatability of a cleaning method for silicon spheres by gravimetric measurements
Autor(in); Institution
Mecke, Michael; 1.1, Masse, PTB-Braunschweig
Borys, Michael; 1.1, Masse, PTB-Braunschweig
Beyer, Edyta; 7.1, Radiometrie mit Synchrotronstrahlung, PTB-Berlin
Quelle/Jahr Special issue for TC3, TC5, TC16 and TC22. Acta IMEKO: 9 (2020), 5, 12 - 16
Availability [online only]
ISSN 2221-870X
DOI
Verlag Braunschweig:
Freie Schlagworte New SI ; kilogram ; silicon spheres ; cleaning method ; repeatability ; mass stability
Zusammenfassung After a short description of the metrological relevance of silicon spheres and the applied cleaning method, mass measurements of silicon spheres in vacuum and the evaluation of the results are outlined. The results show that the mass of silicon spheres in vacuum stabilises in the microgram range within two to threedays after cleaning and that the repeatability of the applied cleaning method can be characterised by a standard deviation in the order of two micrograms.
Kostenfreier Zugang Open Access Gold
Rechteinformation CC BY 3.0 ; Creative Commons Attribution 3.0 License
Themenbereich der Metrologie Masse und abgeleitete Größen

Zitierung

Mecke, M., Borys, M., & Beyer, E. (2020). Investigation of the repeatability of a cleaning method for silicon spheres by gravimetric measurements. Acta IMEKO, 9(5), 12–16. http://doi.org/10.21014/acta_imeko.v9i5.930