Zugriffsnummer 45283
Dokumenttyp Zeitschriftenartikel Open Access Hybrid
Peer Review mit Peer Review
Sprache Englisch
Titel Form deviations caused by lateral displacement errors in annular subaperture stitching interferometry
Autor(in); Institution
Schake, Markus; 4.2, Bild- und Wellenoptik, PTB-Braunschweig
Riebeling, Jörg; Universität Kassel, FB 16 Elektrotechnik/Informatik Messtechnik, Kassel, GERMANY
Ehret, Gerd; 4.2, Bild- und Wellenoptik, PTB-Braunschweig
Quelle/Jahr Optical Engineering: 59 (2020), 12, 1 - 17
Artikelnummer 124105
Availability [online only]
ISSN 0091-3286 (PRINT) ; 1560-2303 (ONLINE)
DOI
Verlag Bellingham, Wash.: SPIE
Freie Schlagworte annular subaperture ; stitching ; interferometry ; traceability ; measurement uncertainty
Zusammenfassung Annular subaperture stitching interferometry (ASSI) is a common approach for the measurement of aspherical surfaces. A common obstacle of ASSI is the occurrence of lateral displacement errors when the sensor or specimen is repositioned between the subaperture measurements. Our contribution focuses on modeling of the statistical displacement errors. A virtual experiment is presented simulating the propagation of the displacement errors through a cumulative and a global stitching algorithm to the retrieved surface form. For the considered experimental setup, the uncertainty in lateral position depends on the positioning uncertainties of the employed motion system and the uncertainty in the absolute distance measurement between the sensor and specimen. The lateral displacement uncertainty is determined experimentally employing a calibratable lateral grating. Thus, it is traceable to the SI unit of the length (meter). The experimental results show that the lateral displacement errors may be modeled by a normal distribution, and the results of the virtual experiment indicate that the statistical lateral displacement errors transfer linear through the stitching procedure and also cause a normal distributed topography error. This enables the assignment of an expanded uncertainty to each individual sample point employing the Zernike polynomial expression of the topography measurement.
Kostenfreier Zugang Open Access Hybrid
Rechteinformation CC BY 4.0 ; Creative Commons Attribution 4.0 License
Themenbereich der Metrologie Länge, dimensionelle Metrologie
Forschungsprojekt DFG Förderung Interferometrischer Zeilensensor Award Nos. LE 992/7-2 and EH 400/4-2
Förderinformationen (1) Förderername: Deutsche Forschungsgemeinschaft (DFG)
Förderer ID: 0000 0001 2096 9829
Förderer ID Typ: ISNI
Titel der Förderung: (LE 992/7-2, EH 400/4-2) Formmessung an Präzisionsbauteilen mit einem dynamisch nachgeführten interferometrischen Zeilensensor

Zitierung

Schake, M., Riebeling, J., & Ehret, G. (2020). Form deviations caused by lateral displacement errors in annular subaperture stitching interferometry. Optical Engineering, 59(12), 1–17. https://doi.org/10.1117/1.oe.59.12.124105