Zugriffsnummer 45137
Dokumenttyp Konferenzartikel
Peer Review mit Peer Review
Sprache Englisch
Titel Linearity measurements on critical noise thermometer components
Autor(in); Institution
Krause, Christian; 7.6, Kryosensorik, PTB-Berlin
Drung, Dietmar; 7.6, Kryosensorik, PTB-Berlin
Quelle/Jahr 2020 Conference on Precision Electromagnetic Measurements (CPEM):(2020), 2 S.
ISSN 0589-1485 (PRINT) ; 2160-0171 (ONLINE)
ISBN 978-1-7281-5899-0 (print) ; 978-1-7281-5898-3 (online)
DOI
Verlag New York, NY: IEEE
Konferenzangaben CPEM 2020 (Conference on Precision Electromagnetic Measurements), Virtual Conference, 24-26, August, 2020, USA
Zusammenfassung A method for measuring the nonlinearity of noise thermometer components was developed that reduces the demands on the test signal generator. The method was applied to characterize the preamplifier and analog-to-digital converter (ADC) of PTB's new noise thermometer. For the complete signal path, gain nonlinearities below about one part in 10 5 were experimentally demonstrated at input rms noise levels between 71 μV and 284 μV. It was further demonstrated that dither can reduce the ADC nonlinearity by at least one order of magnitude.

Zitierung

Krause, C. & Drung, D. (2020). Linearity measurements on critical noise thermometer components. CPEM 2020 (Conference on Precision Electromagnetic Measurements), Virtual Conference, 24-26, August, 2020, USA. https://doi.org/10.1109/cpem49742.2020.9191699