Zugriffsnummer 44910
Dokumenttyp Konferenzartikel
Peer Review unbekannt
Sprache Englisch
Titel Evaluation of the zero-compensation method based on the comparison of two pulse-driven Josephson voltage standards
Autor(in); Institution
Kraus, Marco; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Behr, Ralf; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Kieler, Oliver; 2.4, Quantenelektronik, PTB-Braunschweig
Palafox, Luis; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Herick, Jonas; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Quelle/Jahr 2020 Conference on Precision Electromagnetic Measurements (CPEM):(2020), 286 - 593
ISSN 2160-0171 (ONLINE) ; 0589-1485 (POD)
ISBN 978-1-7281-5898-3 (PDF) ; 978-1-7281-5899-0 (PoD)
DOI
Verlag New York, NY: IEEE
Konferenzangaben Conference on Precision Electromagnetic Measurements (CPEM), Virtual Conference, 24-28, August, 2020
Freie Schlagworte AC Josephson voltage standards ; pulse-driven Josephson arrays ; signal synthesis ; voltage measurements
Zusammenfassung This paper reports on a comparison of two pulsedriven Josephson voltage standards using the zero compensation pulse-bias method. Each system produces a sinusoidal voltage with an RMS signal amplitude of 10 mV for signal frequencies up to 500 kHz. By comparing the results to measurements using the conventical pulse-bias method, we can evaluate the improvements in the absolute accuracy of the synthesized AC signal of each system as well as in the deviation between both systems at signal frequencies above 10 kHz.
Themenbereich der Metrologie Elektrizität und Magnetismus

Zitierung

Kraus, M., Behr, R., Kieler, O., Palafox, L., & Herick, J. (2020). Evaluation of the zero-compensation method based on the comparison of two pulse-driven Josephson voltage standards. Conference on Precision Electromagnetic Measurements (CPEM), Virtual Conference, 24-28, August, 2020. https://doi.org/10.1109/cpem49742.2020.9191757