Zugriffsnummer 44843
Dokumenttyp Zeitschriftenartikel
Peer Review mit Peer Review
Sprache Englisch
Titel Calibration of ultrastable low-noisecurrent amplifiers without direct use of acryogenic current comparator
Autor(in); Institution
Götz, Martin; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Drung, Dietmar; 7.6, Kryosensorik, PTB-Berlin
Krause, Christian; 7.6, Kryosensorik, PTB-Berlin
Mortara, A.; Eidgenössisches Institut für Metrologie (METAS), Bern-Wabern, SWITZERLAND
Quelle/Jahr Metrologia: 57 (2020), 5, 1 - 9
Artikelnummer 055008
ISSN 0026-1394 (print) ; 1681-7575 (online)
DOI
Verlag Bristol: IOP Publishing
Freie Schlagworte amplifiers ; calibration ; electric current measurement ; instrumentation and measurement ; measurement units and standards
Zusammenfassung The established calibration of the transresistance of the ultrastable low-noise current amplifier (ULCA) directly involves a cryogenic current comparator (CCC). Here, we present an alternative approach based on a calibrated set of three resistors and three digital voltmeters. This approach was experimentally verified for a two-channel ULCA with nominally 1 GΩ transresistance using PTB's travelling 12 bit CCC setup in an on-site comparison at METAS. Series of interleaved experiments included calibrations of the input and output stages of two different ULCA channels. The consistency found between the conventional calibration with PTB's CCC and calibration with METAS' set of three resistors-each traced to the quantum Hall resistance with METAS' CCC-confirms that the alternative traceability chain provides valuable results. If the standard resistors are calibrated without using a CCC, the new method even opens the possibility to calibrate the ULCA without any CCC in the traceability chain involved, however, typically at the expense of increased uncertainty.
Themenbereich der Metrologie Elektrizität und Magnetismus
Forschungsprojekt 15SIB08: e-SI-Amp: Quantum realisation of the SI ampere
Förderinformationen (1) Förderername: European Commission (EC)
Förderer ID: 0000 0001 2242 8989
Förderer ID Typ: ISNI
Förderprogramm: EMPIR 2015 SI Broader Scope
Titel der Förderung: 15SIB08: e-SI-Amp: Quantum realisation of the SI ampere
Förderungsnummer: 15SIB08

Zitierung

Götz, M., Drung, D., Krause, C., & Mortara, A. (2020). Calibration of ultrastable low-noisecurrent amplifiers without direct use of acryogenic current comparator. Metrologia, 57(5), 1–9. https://doi.org/10.1088/1681-7575/ab8f7a