Zugriffsnummer 44601
Dokumenttyp Zeitschriftenartikel
Peer Review mit Peer Review
Sprache Englisch
Titel Calibration of thermocouples from 419,527 ºC (freezing point of Zn) up to 1492 ºC (melting point of the Pd-C eutectic), by the temperature fixed point and comparison methods
Autor(in); Institution
Campo, Dolores del; Centro Espanol de Metrologia (CEM), Tres Cantos, SPAIN
Abdelaziz, Yasser A.; National Institute for Standards (NIS), Giza, EGYPT
Anagnostou, Miltiadis; Hellenic Institute of Metrology (EIM), Thessaloniki, GREECE
Arifovic, Narcisa; Turkiye Bilimsel ve Teknologika Arastirma Kurumu (TUBITAK), Gebze/ Kocaeli, TURKEY
Bergerud, Reidun Anita; Justervesenet (JV), Kjeller, NORWAY
Bojkovski, Jovan; University of Ljubljana, Ljubljana, SLOVENIA
Ciocarlan, E.; National Institute of Metrology (INM), Bucuresti, ROMANIA
Edler, Frank; 7.4, Temperatur, PTB-Berlin
Elliot, Claire; National Physical Laboratory (NPL), Teddington, Middlesex, UNITED KINGDOM
Hathela, Ossi; Mittatekniikan Keskus (MIKES), Espoo, FINLAND
Hodzic, Nedzadeta; Institut Metrology Bosnia and Herzegovina (IMBiH), Sarajevo, BOSNIA AND HERZEGOVINA
Holmsten, Magnus; Research Institutes of Sweden (RISE), Boras, SWEDEN
Iacomini, Luigi; Istituto Nazionale di Ricerca Metrologica (INRIM), Torino, ITALY
Kozicki, Marek; Central Office of Measures (GUM), Warsaw, POLAND
Nedialkov, Sasho; Bulgarian Institute of Metrology (BIM), Sofia, BULGARIA
Sadli, Mohamed; Conservatorie National des Arts et Métiers (CNAM), La Plaine-Saint Denis, FRANCE
Simic, Slavica; Directorate of Measures and Precious Metals (DMDM), Belgrade, SERBIA
Strnad, Radek; Czech Metrology Institute (CMI), Brno, CZECH REPUBLIC
Turzó-András, Emese; Magyar Kereskedelmi Engedelyezesi Hivatal (MKEH), Budapest, HUNGARY
Quelle/Jahr Metrologia: 57 (2020), 1A, Tech. Suppl., 121 S.
Availability [online only]
ISSN 1681-7575 (ONLINE)
DOI
Verlag Bristol: IOP
Zusammenfassung This report present the results of the EURAMET inter-comparison carried out to compare the calibration of thermocouples from 419,527 °C (freezing point of Zn) up to 1492 °C (melting point of the Pd-C eutectic). This inter-comparison is intended to be used to support the calibration and measurement capabilities (CMCs) of the participants in the calibration of thermocouples. The comparison was organized in three loops with nineteen participating laboratories and it allowed the performance of the measurement either in fixed points and/or by comparison. The method used to analyse the results of the comparison was the generalized weighted mean that takes into account the full covariance matrix that includes correlations between the participants which have similar traceability sources and between measurements performed by the same laboratory (i.e. the pilots that performed measurements at the beginning and at the end of each loop and the measurements performed by the same laboratory in fixed points and by comparison at the same calibration point).

Zitierung

Campo, D. D., Abdelaziz, Y. A., Anagnostou, M., Arifovic, N., Bergerud, R. A., Bojkovski, J., Ciocarlan, E., Edler, F., Elliot, C., Hathela, O., Hodzic, N., Holmsten, M., Iacomini, L., Kozicki, M., Nedialkov, S., Sadli, M., Simic, S., Strnad, R., & Turzó-András, E. (2020). Calibration of thermocouples from 419,527 ºC (freezing point of Zn) up to 1492 ºC (melting point of the Pd-C eutectic), by the temperature fixed point and comparison methods. Metrologia, 57(1A, Tech. Suppl.), 121 S. https://doi.org/10.1088/0026-1394/57/1a/03006

Exportieren