Zugriffsnummer 44262
Dokumenttyp Zeitschriftenartikel
Peer Review mit Peer Review
Sprache Englisch
Titel Reference dosimetry in MRI-linacs: evaluation of available protocols and data to establish a code of practice
Autor(in); Institution
Pooter, Jacco de; Ionizing Radiation, VSL, Delft, THE NETHERLANDS
Billas, Ilias; National Physical Laboratory (NPL), Teddington, UK
Prez, Leon Armand de; Ionizing Radiation, VSL, Delft, THE NETHERLANDS
Duane, Simon; National Physical Laboratory (NPL), Teddington, UK
Kapsch, Ralf-Peter; 6.2, Dosimetrie für Strahlentherapie und Röntgendiagnostik, PTB-Braunschweig
Karger, Christian; German Cancer Research Center (DKFZ), Division of Medical Physics in Radiation Oncology, Heidelberg, GERMANY
Asselen, Bram van; University Medical Center Utrecht, Department of Radiotherapy, Utrecht, THE NETHERLANDS
Wolthaus, Jochem W. H.; University Medical Center Utrecht, Department of Radiotherapy, Utrecht, THE NETHERLANDS
Quelle/Jahr Physics in Medicine and Biology: 66 (2021), 5, 1 - 38
Artikelnummer 05TR02
ISSN 0031-9155 (PRINT) ; 1361-6560 (ONLINE)
DOI
Verlag Bristol: IOP Publishing
Freie Schlagworte reference dosimetry ; MRI-linacs ; MR-guided radiotherapy ; Monte Carlo ; Ionization chambers ; Codes of Practice ; primary standards
Zusammenfassung With the rapid increase in clinical treatments with MRI-linacs, a consistent, harmonized and sustainable ground for reference dosimetry in MRI-linacs is needed. Specific for reference dosimetry in MRI-linacs is the presence of a strong magnetic field. Therefore, existing Code of Practices (CoPs) are inadequate. In recent years, a vast amount of papers has been published in relation to this topic. The purpose of this review paper is twofold: to give an overview and evaluate the existing literature for reference dosimetry in MRI-linacs and to discuss whether this literature and datasets are adequate and complete to serve as basis for the development of a new or to extend existing CoPs. This review is prefaced with an overview of existing MRI-linac facilities. Then an introduction on the physics of radiation transport in magnetic fields is given. The main part of the review is devoted to the evaluation of the literature with respect to the following subjects: • beam characteristics of MRI-linac facilities; • formalisms for reference dosimetry in MRI-linacs; • characteristics of ionization chambers in the presence of magnetic fields; • ionization chamber beam quality correction factors; • ionization chamber magnetic field correction factors. The review is completed with a discussion whether the existing literature is adequate to serve as basis for a CoP. In addition, it highlights subjects for future research on this topic.
Themenbereich der Metrologie Ionisierende Strahlung

Zitierung

Pooter, J. D., Billas, I., Prez, L. A. D., Duane, S., Kapsch, R.-P., Karger, C., Asselen, B. V., & Wolthaus, J. W. H. (2021). Reference dosimetry in MRI-linacs: evaluation of available protocols and data to establish a code of practice. Physics in Medicine and Biology, 66(5), 1–38. https://doi.org/10.1088/1361-6560/ab9efe