| Zugriffsnummer | 44037 |
| Dokumenttyp | Konferenzartikel in Zeitschrift |
| Peer Review | mit Peer Review |
| Sprache | Englisch |
| Titel | Externally referenced current source with stability down to 1 nA/A at 50 mA |
| Autor(in); Institution |
Fan, Isaac; 8.2, Biosignale, PTB-Berlin
Behr, Ralf; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Drung, Dietmar; 7.6, Kryosensorik, PTB-Berlin
Krause, Christian; 7.2, Kryophysik und Spektrometrie, PTB-Berlin
Götz, Martin; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Pesel, Eckart; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Scherer, Hansjörg; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
|
| Quelle/Jahr | 2018 Conference on Precision Electromagnetic Measurements (CPEM 2018), Full Digest. IEEE Transactions on Instrumentation and Measurement: 68 (2018), 6, 2129 - 2135 |
| ISSN | 0018-9456 (PRINT) ; 1557-9662 (ONLINE) |
| DOI | |
| Verlag | Piscataway, NJ: IEEE |
| Konferenzangaben | Conference on Precision Electromagnetic Measurements ; CPEM 2018, Paris, 08-13, July, 2018, France |
| Freie Schlagworte | Resistors ; Thermal stability ; Circuit stability ; Voltage measurement ; Servomotors ; Current measurement ; Voltage control |
| Zusammenfassung | The noise and stability of a current source are limited by the quality of components used for construction, such as the voltage reference chips, resistors, and power supplies. A state-of-the-art commercial current source typically exhibits a stability no better than parts in 107. In this paper, we present a way to implement a current source with “external” electrical references. This approach allows the current source stability to be decoupled from the noise characteristics of the main power supply circuit. Stability of about a few parts in 109 in a day has been reached at 50 mA, representing the most stable current source at the milliampere range to date. |
| Themenbereich der Metrologie | Metrologie in der Medizin |
Zitierung