Titel International intercomparison of TEPC systems used for radiation protection
Autor(in); Institution
Alberts, Wolfgang G.; 7.3, Neutronendosimetrie, PTB-Braunschweig
Dietz, Ernst; 7.32, Neutronendosimeter, PTB-Braunschweig
Guldbakke, Sigmund; 7.21, Neutronenflussdichte, PTB-Braunschweig
Kluge, Hermann; 7.32, Neutronendosimeter, PTB-Braunschweig
Schuhmacher, Helmut; 7.52, Messtechnik und Prozessdatenverarbeitung, PTB-Braunschweig
Quelle/Jahr Radiation Protection Dosimetry: 29 (1989), 1/2, 47 - 53
ISSN 0144-8420
DOI
Verlag Oxford: Oxford University Press
Zusammenfassung The irradiation conditions and procedures of an international intercomparison of tissue-equivalent proportional counters used for radiation protection measurements are described. The irradiations took place in monoenergetic reference neutron fields produced at the research reactor and accelerator facilities of the PTB, Braunschweig, in the energy range from thermal neutrons to 14.8 MeV. In addition, measurements were performed in 60Co and D2O-moderated 252Cf radiation fields. Prototype instruments from seven European groups were investigated. The results of the measurements are summarised and briefly discussed. Auch enthalten in: PTB 1975-1995 : Präsidentschaft Dieter Kind, 1995 [Aufsatzsammlung]

Zitierung

Alberts, W. G., Dietz, E., Guldbakke, S., Kluge, H., & Schuhmacher, H. (1989). International intercomparison of TEPC systems used for radiation protection. Radiation Protection Dosimetry, 29(1/2), 47–53. https://doi.org/10.1093/oxfordjournals.rpd.a080529