Zugriffsnummer 43982
Dokumenttyp Zeitschriftenartikel Open Access Hybrid
Peer Review mit Peer Review
Sprache Englisch
Titel Correction factors for two new reference beta radiation fields
Autor(in); Institution
Behrens, Rolf; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Quelle/Jahr Metrologia: 57 (2020), 6, 1 - 15
Artikelnummer 065005
ISSN 0026-1394 (PRINT) ; 1681-7575 (ONLINE)
DOI
Verlag Bristol: IOP Publishing
Klassifikationscode PACS: 87.53.Bn ; PACS: 29.40.Cs ; PACS: 06.20.F ; PACS: 07.05.Tp
Freie Schlagworte PTB high performance compute cluster
Zusammenfassung Several correction factors for two new reference beta-particle radiation fields complementing ISO 6980 have been measured and calculated for both primary beta dosimetry and the operation-al quantities in radiation protection. The following correction factors for primary beta dosimetry for the determination of the absorbed dose to tissue, Dt, have been determined by calculations: kba for backscatter from the collecting electrode, kpe for perturbation by the chamber’s side walls, kih for in-homogeneity inside the collecting volume, kSta for the stopping power ratio at different phantom depths, and kSA for the use of the Spencer-Attix theory, while the following have been measured: kbr for the effect of bremsstrahlung and kabs for variations in the attenuation and scattering of beta particles between the source and the collecting volume due to variations from reference conditions and for  differences of the entrance window to a tissue equivalent thickness of 0.07 mm. Furthermore, calcula-tions were undertaken to determine the following correction factors to assess the operational quantities Hp(0.07), Hp(3), H'(0.07), and H'(3): krod(0.07) for the rod instead of the slab phantom, kcyl(3) for the cylinder instead of the slab phantom, as well as k'(0.07) and k'(3) for the ICRU sphere instead of the slab phantom, while the correction factor for oblique radiation incidence has been measured. The newly determined correction factors are ready to be implemented in an updated version of the ISO 6980 series and in the software of the Beta Secondary Standard BSS 2. This article is corrected by 2026 Metrologia 63 019501
Relation
https://doi.org/10.1088/1681-7575/ae3c19 – (Corrigendum (2026))
Kostenfreier Zugang Open Access Hybrid
Rechteinformation CC BY 4.0 ; Creative Commons Attribution 4.0 License
Themenbereich der Metrologie Ionisierende Strahlung

Zitierung

Behrens, R. (2020). Correction factors for two new reference beta radiation fields. Metrologia, 57(6), 1–15. https://doi.org/10.1088/1681-7575/ab8829

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag