Zugriffsnummer 43846
Dokumenttyp Konferenzartikel
Peer Review unbekannt
Sprache Englisch
Titel The fourth primary cesium atomic clock of the PTB
Autor(in); Institution
Bauch, Andreas; 4.41, Zeiteinheit, PTB-Braunschweig
Fischer, Bernd; 4.41, Zeiteinheit, PTB-Braunschweig
Heindorff, Thomas; 4.41, Zeiteinheit, PTB-Braunschweig
Schröder, Roland; 4.41, Zeiteinheit, PTB-Braunschweig
Quelle/Jahr 7th European Frequency and Time Forum : Neuchâtel, Switzerland, March 16-18, 1993:(1993), 531 - 535
Verlag Neuchâtel: FSRM
Konferenzangaben 7th European Frequency and Time Forum Neuchatel, Neuchâtel, 16-18, March, 1993, Switzerland
Freie Schlagworte cesium atomic clock ; realization of SI second
Zusammenfassung The CS3, the second primary atomic clock with vertical cesium beams, has been put into continuous operation in autumn 1992. Its construction and its overall characteristics are identical with those of CS4. In the first 150 days of operation the CS3 frequency shifts related to the C-field, the second-order Doppler effect and the cavity phase difference, and their temporal stability were measured. In the period of common operation the relative frequencies of CS3 and of CS4 agree within (25±9) · 10-15. The measured frequency instability of CS3 is 8,6·10-12 √s/τ for averaging times τ, 100 s ≤ τ ≤ 20000 s.

Zitierung

Bauch, A., Fischer, B., Heindorff, T., & Schröder, R. (1993). The fourth primary cesium atomic clock of the PTB. 7th European Frequency and Time Forum Neuchatel, Neuchâtel, 16-18, March, 1993, Switzerland.