| Zugriffsnummer | 41828 |
| Dokumenttyp | Zeitschriftenartikel |
| Peer Review | mit Peer Review |
| Sprache | Englisch |
| Titel | Final report: key comparison EURAMET.AUV.A-K5 |
| Autor(in); Institution |
Avison, Janine; National Physical Laboratory (NPL), Teddington, UK
Harris, Peter; National Physical Laboratory (NPL), Teddington, UK
Robinson, Stephen; National Physical Laboratory (NPL), Teddington, UK
Rodrigues, Dominique; Laboratoire National de Métrologie et d’Essais (LNE), Trappes, FRANCE
Guglielmone, Claudio; Istituto Nazionale di Ricerca Metrologica (INRiM), Torino, ITALY
Hof, Christian; Swiss Federal Office of Metrology (METAS), Bern-Wabern, SWITZERLAND
Sadikoglu, Enver; UME, TUBITAK, TURKEY
Chushkov, Marin; BIM, Sofia, BULGARIA
Zelenika, Slobodan; Directorate of Measures and Precious Metals (DMDM), Belgrade, SERBIA
Rosenkrantz, Peter; BEV, Wien, AUSTRIA
Andersson, Håkan; RISE - Research Institutes of Sweden, Boras, SWEDEN
Fedtke, Thomas; 1.6, Schall, PTB-Braunschweig
Koch, Christian; 1.6, Schall, PTB-Braunschweig
Martin, Nieves Medina; Centre Espanol de Metrologia (CEM), Madrid, SPAIN
Shawky, Hany; NIS, National Institute for Standards, El-Giza, EGYPT
|
| Quelle/Jahr | Metrologia: 56 (2019), 1A, Tech. Suppl., 1 - 39 |
| Artikelnummer | 09001 |
| ISSN | 0026-1394 (PRINT) ; 1681-7575 (ONLINE) |
| DOI | |
| URL | |
| Verlag | Bristol: IOP Publishing |
| Freie Schlagworte | acoustics, microphones ; metrology ; key comparison ; pressure calibration |
| Zusammenfassung | This is the final (Draft B) report for key comparison EURAMET.AUV.A-K5 on the pressure calibration of laboratory standard microphones in the frequency range from 2 Hz to 10 kHz. Twelve national metrology institutes took part in the key comparison and the National Physical Laboratory piloted the project. Two travelling standard microphones were circulated to the participants and results in the form of regular calibration certificates were collected throughout the project. The analysis used the results for one of the microphones only and values for both sensitivity level and sensitivity phase and have been linked to the CCAUV.A K5 key comparison reference value (KCRV) via two linking laboratories (NPL and INRIM). The main text of this paper is published as: NPL report AC 18 (2018) |
| Themenbereich der Metrologie | Akustik, Ultraschall, Beschleunigung |
Zitierung
Avison, J., Harris, P., Robinson, S., Rodrigues, D., Guglielmone, C., Hof, C., Sadikoglu, E., Chushkov, M., Zelenika, S., Rosenkrantz, P., Andersson, H., Fedtke, T., Koch, C., Martin, N. M., & Shawky, H. (2019). Final report: key comparison EURAMET.AUV.A-K5. Metrologia, 56(1A, Tech. Suppl.), 1–39. https://doi.org/10.1088/0026-1394/56/1a/09001