Titel Determining the dose rate dependence of different active personal dosemeters in standardized pulsed and continuous radiation fields
Autor(in); Institution
Hupe, Oliver; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Friedrich, Sabrina; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Vanhavere, Filip; SCK-CEN, Mol, BELGIUM
Brodecki, Marcin; Nofer Institute of Occupational Medicine (NIOM), Lodz, POLAND
Quelle/Jahr Radiation Protection Dosimetry:(2019), 1 - 8
ISSN 0144-8420 (PRINT) ; 1742-3406 (ONLINE)
DOI
Verlag Oxford: Oxford University Press
Zusammenfassung In 2016 the EURADOS working group 12 started a project to gather reliable data concerning the behavior of active personal dosemeters used by European hospitals in standardized pulsed and hospital fields. The project is devided in 6 work packages. This paper concentrates on the work packages 2 and 3 which deal amongst others with the dose rate dependence of the response of the active personal dosemeters in laboratory tests. The results from these tests represent the basis for the following tests in realistic irradiation conditions to those in hospitals. The dependence on the dose rate was determined for 10 dosemeter types with continuous radiation of a 60Co source and pulsed radiation of the reference radiation facility of PTB. Furthermore irradiations were done with continuous radiation for a 137Cs source of the KIT and with continuous X-ray radiation selected RQR radiation quality (RQR 5 and RQR 9) of NIOM.
Themenbereich der Metrologie Ionisierende Strahlung

Zitierung

Hupe, O., Friedrich, S., Vanhavere, F., & Brodecki, M. (2019). Determining the dose rate dependence of different active personal dosemeters in standardized pulsed and continuous radiation fields. Radiation Protection Dosimetry, 1–8. https://doi.org/10.1093/rpd/ncz173

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag