Zugriffsnummer 41357
Dokumenttyp Konferenzartikel
Peer Review unbekannt
Sprache Englisch
Titel Purpose-built air bath for use with NMIA calculable cross-capacitor measurement chain
Autor(in); Institution
Johnson, Heather; National Measurement Institute (NMI), Lindfield, NSW, AUSTRALIA
Coogan, Peter; National Measurement Institute (NMI), Lindfield, NSW, AUSTRALIA
Small, Greig; National Measurement Institute (NMI), Lindfield, NSW, AUSTRALIA
Palafox Gamir, Luis; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Quelle/Jahr 2018 Conference on Precision Electromagnetic Measurements (CPEM 2018), Full Digest:(2018), 742 - 743
ISSN 2160-0171 (ELECTRONIC) ; 0589-1485 (POD)
ISBN 978-1-5386-0974-3 (PDF) ; 978-1-5386-0973-6 (USB) ; 978-1-5386-2130-1 (PoD)
DOI
Verlag Piscataway, NJ: IEEE
Konferenzangaben Conference on Precision Electromagnetic Measurements ; CPEM 2018, Paris, 08-13, July, 2018, France
Freie Schlagworte Capacitance measurement ; measurement standards ; precision measurements ; temperature control
Zusammenfassung This paper describes the design of an air bath for maintaining the reference capacitors associated with the new NMIA calculable cross-capacitor measurement system. The air bath is designed to maintain a set temperature in the range 15 °C to 25 °C with a stability of 0.1 mK over the short time periods during which the capacitors are measured.
Themenbereich der Metrologie Elektrizität und Magnetismus