| Zugriffsnummer | 40931 |
| Dokumenttyp | Konferenzartikel in Zeitschrift |
| Peer Review | mit Peer Review |
| Sprache | Englisch |
| Titel | Influence of a magnetic field on the dose deposited by a 6MV linac at tissue interfaces |
| Autor(in); Institution |
Richter, Sebastian; University Hospital and Medical Faculty Eberhard Karls University Tübingen, Department of Radiation Oncology, Section for Biomedical Physics, Tübingen, GERMANY
Mönnich, David; University Hospital and Medical Faculty Eberhard Karls University Tübingen, Department of Radiation Oncology, Section for Biomedical Physics, Tübingen, GERMANY
Pojtinger, Stefan; 6.2, Dosimetrie für Strahlentherapie und Röntgendiagnostik, PTB-Braunschweig; University Hospital and Medical Faculty Eberhard Karls University Tübingen, Department of Radiation Oncology, Section for Biomedical Physics, Tübingen, GERMANY
Dohm, Oliver; University Hospital and Medical Faculty Eberhard Karls University Tübingen, Department of Radiation Oncology, Division for Medical Physics, Tübingen, GERMANY
Zips, Daniel; University Hospital and Medical Faculty Eberhard Karls University Tübingen, Department of Radiation Oncology, Tübingen, GERMANY; German Cancer Research Center DKFZ, German Cancer Consortium DKTK partner site Tübingen, Heidelberg, GERMANY
Thorwarth, Daniela; University Hospital and Medical Faculty Eberhard Karls University Tübingen, Department of Radiation Oncology, Section for Biomedical Physics, Tübingen, GERMANY
|
| Quelle/Jahr | Radiotherapy and Oncology: 127 (2018), Suppl. 1, 469 - 470 |
| Artikelnummer | PO-0887 |
| ISSN | 0167-8140 (print) ; 1879-0887 (online) |
| DOI | |
| Verlag | Amsterdam: Elsevier |
| Konferenzangaben | 37th Annual Meeting of the European Society of Radiotherapy and Oncology, Barcelona, 20-24, April, 2018, Spain |
| Freie Schlagworte | Dosimetrie ; MR-Linac ; EGSnrc |
| Zusammenfassung | A MR-linac provides high soft-tissue contrast imaging during radiotherapy treatments: magnetic resonance guided radiotherapy (MRgRT). However, trajectories of secondary electrons are influenced by the magnetic field and may deposit dose outside of the primary beam. This work investigated changes of dose depositions at water-air and tissue-lung interfaces due to a magnetic field (B-field) transverse to the beam direction by means of Monte Carlo (MC) simulations and opposing beams setups. |
| Kostenfreier Zugang | Open Access Hybrid |
| Rechteinformation | CC BY-NC-ND 4.0 ; Creative Commons Attribution NonCommercial NoDerivatives 4.0 License |
| Themenbereich der Metrologie | Ionisierende Strahlung |
| Forschungsprojekt | 15HLT08: MRgRT: Metrology for MR guided radiotherapy |
| Förderinformationen (1) |
Förderername: European Commission (EC)
Förderer ID: 0000 0001 2242 8989 Förderer ID Typ: ISNI Förderprogramm: EMPIR 2015 Health Titel der Förderung: 15HLT08: MRgRT: Metrology for MR guided radiotherapy Förderungsnummer: 15HLT08 |
Zitierung
Richter, S., Mönnich, D., Pojtinger, S., Dohm, O., Zips, D., & Thorwarth, D. (2018). Influence of a magnetic field on the dose deposited by a 6MV linac at tissue interfaces. Radiotherapy and Oncology, 127(Suppl. 1), 469–470. https://doi.org/10.1016/s0167-8140(18)31197-6