Titel Report of IAEA - PTB bilaterial comparison: Reference airkerma rate for high dose rate 192lr and 60Co brachytherapy sources
Autor(in); Institution
Czap, Ladislav; IAEA, Vienna, AUSTRIA
Bokulic, T.; IAEA, Vienna, AUSTRIA
Toroi, P.; IAEA, Vienna, AUSTRIA
Behrens, Rolf; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Grote, Friedrike; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Kasper, Angelika; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Quelle/Jahr SSDL Newsletter: 71 (2020), 7 - 10
ISSN 1001-2699
Verlag Wien: International Atomic Energy Agency (IAEA)
Zusammenfassung A comparison of the standards for reference airkerma rate of the Physikalisch-Technische Bundesanstalt (PTB) and of the International Energy Agency (IAEA) was carried out with 192Ir and 60Co high dose rate brachytherapy sources in May 2018. Comparison results were calculated based on the calibration coefficients for the transfer standard and expressed as the ratio of the IAEA and the PTB values. The comparison results were 1.0022 (uc=0.0065) and 1.0022 (uc = 0,0065), k = 1, for 192Ir and 60Co sources, respectively.
Themenbereich der Metrologie Ionisierende Strahlung

Zitierung

Czap, L., Bokulic, T., Toroi, P., Behrens, R., Grote, F., & Kasper, A. (2020). Report of IAEA - PTB bilaterial comparison: Reference airkerma rate for high dose rate 192lr and 60Co brachytherapy sources. SSDL Newsletter, 71, 7–10.