| Titel | Experiences with cryopumps in calibration devices |
| Autor(in); Institution |
Grosse, Gottfried; 7.23, Vakuumphysik, PTB-Berlin
Messer, Günter; 7.23, Vakuumphysik, PTB-Berlin
|
| Quelle/Jahr | Vacuum: 20 (1970), 9, 373 - 376 |
| ISSN | 0042-207X |
| DOI | |
| Verlag | Amsterdam: Elsevier |
| Zusammenfassung | A room-temperature calibration system is described in which calibrating pressures between 10P9 and 105 torr are maintained by a helium evaporator cryopump with a liquid helium consumption of 2–3 litres/hour. The major advantage of this system is that, due to the high pumping speed of the cryopump, an orifice of relatively large conductance (100 litres/sec) may be used. Hydrogen background pressures of 1010 torr may be maintained for a sufficient time to permit the use of all test gases except helium and hydrogen. |