Autor(in); Institution
Grosse, Gottfried; 7.23, Vakuumphysik, PTB-Berlin
Messer, Günter; 7.23, Vakuumphysik, PTB-Berlin
Quelle/Jahr Vacuum: 20 (1970), 9, 373 - 376
ISSN 0042-207X
DOI
Verlag Amsterdam: Elsevier
Zusammenfassung A room-temperature calibration system is described in which calibrating pressures between 10P9 and 105 torr are maintained by a helium evaporator cryopump with a liquid helium consumption of 2–3 litres/hour. The major advantage of this system is that, due to the high pumping speed of the cryopump, an orifice of relatively large conductance (100 litres/sec) may be used. Hydrogen background pressures of 1010 torr may be maintained for a sufficient time to permit the use of all test gases except helium and hydrogen.
Themenbereich der Metrologie Thermometrie

Zitierung

Grosse, G. & Messer, G. (1970). Experiences with cryopumps in calibration devices. Vacuum, 20(9), 373–376. https://doi.org/10.1016/s0042-207x(70)80036-7