Titel EURAMET project No. 1225: intercomparison of very low air speed standard facilities (0.05 m/s to 1 m/s)
Autor(in); Institution
Pachinger, Dietmar; BEV/E+E, Designated Laboratory, Engerwitzdorf, AUSTRIA
De Huu, Marc; METAS, Physics and Chemistry - Mechanical Quantities and Ionising Radiation - Flow and Hydrometry, Bern-Wabern, SWITZERLAND
Müller, Harald; 1.4, Gase, PTB-Braunschweig
Care, Isabelle; LNE-CETIAT, Laboratoire Anemometrie, Domaine scientifique de la Doua, Villeurbanne, FRANCE
Frederiksen, John; Danish Technological Institute, Installation and Calibration Technology, Teknologiparken, Aarhus C, DENMARK
Piccato, Aline; Istituto Nazionale di Ricerca Metrologica (INRiM), Torino, ITALY
Bertasiene, Agne; LEI, Heat Equipment Research and Testing Laboratory, Kaunas, LITHUANIA
Quelle/Jahr Metrologia: 54 (2017), 1A, Tech. Suppl., 1 - 22
Artikelnummer 07021
Availability [online only]
ISSN 0026-1394 (PRINT) ; 1681-7575 (ONLINE)
DOI
Verlag Bristol: IOP
Zusammenfassung For the first time a EURAMET comparison of standard facilities in the field of low air speed is presented. The air velocities range from 0.05 m/s up to 1 m/s. Two different thermal anemometers are used as Transfer Standards (TS), namely a ball-type and a planar-type anemometer. Due to their construction it is possible to analyze different components of the flow vector. Orientation-dependent measurements which are investigated in addition might also deliver information concerning the components of the flow vector. A Comparison Reference Value (CRV) is calculated with the help of the chi-squared test according to the procedure A presented by Cox and the degree of equivalence of the laboratory data to the CRV is determined. Both TS show reasonable results and a high degree of equivalence. Furthermore the degree of equivalence of the laboratory data to each other is investigated in detail. Orig. title of the final report: EURAMET.M.FF-S9: Intercomparison of very low air speed standard facilities (0.05 - 1 m/s)

Zitierung

Pachinger, D., De Huu, M., Müller, H., Care, I., Frederiksen, J., Piccato, A., & Bertasiene, A. (2017). EURAMET project No. 1225: intercomparison of very low air speed standard facilities (0.05 m/s to 1 m/s). Metrologia, 54(1A, Tech. Suppl.), 1–22. https://doi.org/10.1088/0026-1394/54/1a/07021