Zugriffsnummer 38422
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel Traceability of bridge amplifiers in tared measurement applications
Autor(in); Institution
Brüge, Andreas; 1.2, Festkörpermechanik, PTB-Braunschweig
Quelle/Jahr IMEKO TC3 Conference 2017: Proceedings:(2017), 5 S.
Availability [online only]
URL
Verlag IMEKO
Konferenzangaben IMEKO TC3, TC5 and TC22 International Conference 2017, Helsinki, 30, May - 01, June, 2017, Finland
Freie Schlagworte voltage ratio ; traceability ; bridge amplifier ; uncertainty ; linearity
Zusammenfassung The influence of the primary voltage ratio standard uncertainty on strain gauge measurements, especially at secondary levels like torque wrench calibration, is discussed. A method is presented to overcome its 1/x behaviour, which affects the uncertainty budget at lower ratios for the usual point-by-point transfer of uncertainty. Monte Carlo simulations of fitted correction values reduce the relative expanded contribution of the traceability uncertainty to less than 1·10-5 for tared measurements with bridge amplifiers in comparison to five times more for point-by-point traceability. Considering typical further uncertainty contributions, a tared measurement of voltage ratios is possible with a combined relative expanded uncertainty of less than 1.8·10-5.

Zitierung

Brüge, A. (2017). Traceability of bridge amplifiers in tared measurement applications. IMEKO TC3, TC5 and TC22 International Conference 2017, Helsinki, 30, May - 01, June, 2017, Finland.

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag