Zugriffsnummer 37844
Dokumenttyp Zeitschriftenartikel
Peer Review unbekannt
Sprache Englisch
Titel Correction factors for attenuation and scattering of the wall of a cylindrical ionization chamber in the 6 MeV to 7 MeV high-energy photon reference field
Autor(in); Institution
Kowatari, M.; Japan Atomic Energy Agency (JAEA), Nuclear Science Research Institute, Department of Radiation Protection, Ibaraki, JAPAN
Zutz, Hayo; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Hupe, Oliver; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Quelle/Jahr Radiation Protection Dosimetry: 178 (2018), 1, 48 - 56
ISSN 0144-8420 (PRINT) ; 1742-3406 (ONLINE)
DOI
Verlag Oxford: Oxford University Press
Zusammenfassung In high enery photon reference fields the value fo the air kerma rate is determined by using ionization chambers. From the charge collected inside the ionization chamber the dose can be calculated using a set of calibration and correction factors according to ISO 4037-2. A crucial parameter is the correction for the attenuation and scattering of the Primary Radiation due to the chamber wall. This parameter can be determined using Monte Carlo calculations. The evaluation of the factor was performed for a commercially available ionization chamber of the type Victoreen 550-3 under different build-up conditions. The results were verified by measurements in the R-F high energy photon fields according to ISO 4037-1 at the Physikalisch-Technische Bundesanstalt (PTB) and the Japan Atomic Energy Agency (JAEA).

Zitierung

Kowatari, M., Zutz, H., & Hupe, O. (2018). Correction factors for attenuation and scattering of the wall of a cylindrical ionization chamber in the 6 MeV to 7 MeV high-energy photon reference field. Radiation Protection Dosimetry, 178(1), 48–56. https://doi.org/10.1093/rpd/ncx073

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag