Zugriffsnummer 37814
Dokumenttyp Zeitschriftenartikel
Sprache Englisch
Titel Direct comparison of cryogenic radiometers from the BIPM and the PTB
Autor(in); Institution
Goebel, R.; Bureau International des Poids et Mesures, Sèvres Cedex, FRANCE
Pello, R.; Bureau International des Poids et Mesures, Sèvres Cedex, FRANCE
Stock, Klaus; 4.11, Radiometrische Einheiten, PTB-Braunschweig
Hofer, Helmut; 4.11, Radiometrische Einheiten, PTB-Braunschweig
Quelle/Jahr Metrologia: 34 (1997), 3, 257 - 259
ISSN 0026-1394
DOI
Verlag Sèvres: BIPM
Zusammenfassung For the first time, the cryogenic radiometer of the Bureau International des Poids et Mesures (BIPM) has been taken to a national metrology institute for direct comparison with another cryogenic radiometer. This paper describes a comparison at the Physikalisch-Technische Bundesanstalt (PTB, Braunschweig, Germany) with the PTB cryogenic radiometer. The two systems have been shown to differ, in relative terms, by 2,3 × 10-4 in the measurement of optical power at λ = 632,8 nm with a power level of about 700 μW. The standard uncertainty in the determination of this relative difference is estimated to be 5,7 × 10-5. As the source of this discrepancy, which is larger than the uncertainties, has not yet been found, a closer investigation of the two systems is necessary.

Zitierung

Goebel, R., Pello, R., Stock, K., & Hofer, H. (1997). Direct comparison of cryogenic radiometers from the BIPM and the PTB. Metrologia, 34(3), 257–259. https://doi.org/10.1088/0026-1394/34/3/8

Exportieren