Zugriffsnummer 37671
Dokumenttyp Konferenzartikel in Zeitschrift
Sprache Englisch
Titel A new calibration transformer and measurement setup for bridge standard calibrations up to 5 kHz
Autor(in); Institution
Beug, Marc Florian; 2.1, Gleichstrom und Niederfrequenz, PTB-Braunschweig
Moser, Harald; 2.1, Gleichstrom und Niederfrequenz, PTB-Braunschweig
Kölling, Axel; 2.1, Gleichstrom und Niederfrequenz, PTB-Braunschweig
Quelle/Jahr IEEE Transactions on Instrumentation and Measurement: 66 (2017), 6, 1531 - 1538
ISSN 0018-9456 (PRINT) ; 1557-9662 (ONLINE)
DOI
Verlag New York, NY: IEEE
Konferenzangaben Conference on Precision Electromagnetic Measurements (CPEM) 2016, Ottawa, 10-15, July, 2016, Canada
Freie Schlagworte Static bridge standard calibration ; carrier frequency supply voltage ; inductive voltage divider
Zusammenfassung This paper presents an improved calibration setup for static bridge standards with a carrier frequency supply voltage in the frequency range between 225 Hz and 5 kHz. In the past, the small bridge differential voltage in the mV/V range was calibrated in two subsequent voltage ratio calibrations. However, this procedure is time-consuming and only works reliably if both voltages are stable over time. For this purpose, a new setup was developed that is able to measure the differential voltage of the bridge standard in one single measurement. This reduces the calibration time by approximately two thirds, thereby also reducing the uncertainty.

Zitierung

Beug, M. F., Moser, H., & Kölling, A. (2017). A new calibration transformer and measurement setup for bridge standard calibrations up to 5 kHz. IEEE Transactions on Instrumentation and Measurement, 66(6), 1531–1538. https://doi.org/10.1109/tim.2017.2653520