| Zugriffsnummer | 37503 |
| Dokumenttyp | Zeitschriftenartikel |
| Peer Review | unbekannt |
| Sprache | Englisch |
| Titel | Final report of CCAUV.V-K3: key comparison in the field of acceleration on the complex charge sensitivity |
| Autor(in); Institution |
Sun, Qiao; National Institute of Metrology (NIM), Beijing, CHINA
Yang, Lifeng; National Institute of Metrology (NIM), Beijing, CHINA
Bartoli, Claire; Laboratoire national de métrologie et d'essais (LNE), FRANCE
Veldman, Ian; National Metrology Institute of South Africa (NMISA), SOUTH AFRICA
Ripper, Gustavo P; Instituto Nacional de Metrologia, Qualidade e Tecnologia (INMETRO), Duque de Caxias, BRAZIL
Bruns, Thomas; 1.7, Akustik und Dynamik, PTB-Braunschweig
Licht, Torben Rask; Brüel & Kjaer Sound & Vibration Measurement Ltd - Danish Primary Laboratory for Acoustics Section, DENMARK
Kolasa, Joanna; Glowny Urzad Miar (GUM), Warszawa, POLAND
Hof, Christian; Swiss Federal Office of Metrology (METAS), Bern-Wabern, SWITZERLAND
Pineda, Guillermo Silva; Centro Nacional de Metrologia (CENAM), Queretaro, MEXICO
Dickinson, Laurence; National Measurement Institute of Australia (NMIA), AUSTRALIA
Ota, Akihiro; National Metrology Institute of Japan (NMIJ), Tsukuba, Ibaraki, JAPAN
Cheung, Wan Sup; Korea Research Institute of Standards and Science (KRISS), REPUBLIC OF KOREA
Yankovsky, Alexander; D.I. Mendeleyev Institute for Metrology (VNIIM), St. Petersburg, RUSSIA
Shan, Cui; National Metrology Centre, Agency for Science, Technology and Research Citation (NMC) A*STAR, SINGAPORE
|
| Quelle/Jahr | Metrologia: 54 (2017), 1A, Tech. Suppl., 1 - 79 |
| Artikelnummer | 09001 |
| Availability | [online only] |
| ISSN | 0026-1394 (PRINT) ; 1681-7575 (ONLINE) |
| DOI | |
| Verlag | Bristol: IOP |
| Zusammenfassung | This is the final report for CIPM key comparison CCAUV.V-K3 in the area of 'vibration' (quantity of acceleration). The aim of this comparison was to measure the voltage sensitivity of one accelerometer standard set with primary means at 27 frequencies from 0.1 Hz to 40 Hz. Fourteen Metrology Institutes from five RMOs have participated in the comparison with National Institute of Metrology, P.R. China as pilot lab and Laboratoire National de Métrologie et d'Essais and National Metrology Institute of South Africa as co-pilot labs. One quartz-flexure servo accelerometer of single-ended type and a signal conditioner was circulated among the participants. All but one of the participating laboratories provided their calibration results, which were mostly consistent within their declared expanded uncertainties for magnitude results. Only two participants failed to contribute to the KCRV values calculated for five frequencies. For phase shift, three participants could not contribute to the calculation of the KCRV values in a total of sixteen frequencies. This first low-frequency vibration key comparison revealed the current calibration capabilities of the fourteen participants of five RMOs. |
Zitierung
Sun, Q., Yang, L., Bartoli, C., Veldman, I., Ripper, G. P., Bruns, T., Licht, T. R., Kolasa, J., Hof, C., Pineda, G. S., Dickinson, L., Ota, A., Cheung, W. S., Yankovsky, A., & Shan, C. (2017). Final report of CCAUV.V-K3: key comparison in the field of acceleration on the complex charge sensitivity. Metrologia, 54(1A, Tech. Suppl.), 1–79. https://doi.org/10.1088/0026-1394/54/1a/09001