Titel Efficiency calibration of semiconductor detectors by primary standard sources and Monte Carlo calculations
Autor(in); Institution
Debertin, Klaus; 6.31, Radioaktive Standards, PTB-Braunschweig
Grosswendt, Bernd; 6.12, Strahlungstransportgrößen, PTB-Braunschweig
Quelle/Jahr Nuclear Instruments and Methods in Physics Research: 203 (1982), 1-3, 343 - 352
ISSN 0029-554X
DOI
Verlag Amsterdam: North-Holland
Zusammenfassung The full energy peak efficiency of two Ge(Li), two high-purity germanium and one Si(Li) detector was determined experimentally by using primary standard sources emitting gamma-radiation in the energy range from 5 keV to 3 MeV. In addition, the efficiency was calculated by means of Monte Carlo methods. Measurements and calculations are in excellent agreement for some detectors but disagree for others. Reasons for the deviations are discussed. Monte Carlo calculations cannot replace experimental efficiency determinations but it is demonstrated that they are a useful aid for a reliable interpolation and extrapolation of experimental efficiency data.

Zitierung

Debertin, K. & Grosswendt, B. (1982). Efficiency calibration of semiconductor detectors by primary standard sources and Monte Carlo calculations. Nuclear Instruments and Methods in Physics Research, 203(1-3), 343–352. https://doi.org/10.1016/0167-5087(82)90646-9