Titel Comparison of high temperature and high Reynolds number water flows between PTB and NMIJ
Autor(in); Institution
Furuichi, N.; NMIJ, Tsukuba, JAPAN
Cordova Murillo, Moritz Leopoldo; 7.5, Wärme und Vakuum, PTB-Berlin
Lederer, Thomas; 7.5, Wärme und Vakuum, PTB-Berlin
Terao, Y.; NMIJ, Tsukuba, JAPAN
Quelle/Jahr Flow Measurement and Instrumentation: 52 (2016), 12, 157 - 162
ISSN 0955-5986
DOI
Verlag Amsterdam [u.a.]: Elsevier
Freie Schlagworte Bi-comparison ; Water flowrate ; High temperature ; High Reynolds number ; Differential pressure meter
Zusammenfassung Bi-comparison for high temperature and high Reynolds number water flows was performed between Physikalische-Technische Bundesanstalt (PTB) and National Metrology Institute of Japan (NMIJ) using differential pressure meters. The flowrate range considered herein was from 250 m3/h to 740 m3/h and the temperature range was from 20 °C to 80 °C. The Reynolds number based on the pipe diameter ranges from 4.4×105 to 3.6×106. The expanded uncertainty of volumetric flowrate with a coverage factor k=2 is 0.040% for PTB and 0.070% for NMIJ. The differential pressure meters as the transfer meter were a throat-tapped flow nozzle and an orifice. The results for both flow meters were in good agreement between the two labs. The difference in the average discharge coefficient at the same Reynolds number ranged from −0.05% to 0.03% for the flow nozzle and from −0.01% to 0.05% for the orifice. The results of the comparison indicate a high level of consistency between PTB and NMIJ for high temperature and high Reynolds number water flows.

Zitierung

Furuichi, N., Cordova Murillo, M. L., Lederer, T., & Terao, Y. (2016). Comparison of high temperature and high Reynolds number water flows between PTB and NMIJ. Flow Measurement and Instrumentation, 52(12), 157–162. https://doi.org/10.1016/j.flowmeasinst.2016.10.003

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag