| Zugriffsnummer | 36587 |
| Dokumenttyp | Konferenzartikel |
| Peer Review | unbekannt |
| Sprache | Englisch |
| Titel | Dry cooling of a 10 V programmable Josephson voltage standard array |
| Autor(in); Institution |
Starkloff, Michael; Supracon AG, Jena, GERMANY
Schubert, Marco; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig; Supracon AG, Jena, GERMANY
Lee, Jinni; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Behr, Ralf; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Palafox Gamir, Luis; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Böck, Andreas C.; esz AG calibration + metrology, Eichenau, GERMANY
Fleischmann, Philip M.; esz AG calibration + metrology, Eichenau, GERMANY
Peiselt, K.; Leibniz Institute of Photonic Technology, Jena, GERMANY
|
| Quelle/Jahr | CPEM 2016, Conference on Precision Electromagnetic Measurements: conference digest:(2016), 2 S. |
| Availability | [USB-Stick] ; file name: 023.pdf |
| ISBN | 978-1-4673-9134-4 ; 978-1-4673-9133-7 |
| DOI | |
| Verlag | [s.l.]: IEEE |
| Konferenzangaben | Conference on Precision Electromagnetic Measurements (CPEM) 2016, Ottawa, 10-15, July, 2016, Canada |
| Freie Schlagworte | AC Quantum Voltmeter ; cooling ; cryocooler ; Josephson junctions ; programmable Josephson voltage standard |
| Zusammenfassung | This paper describes the properties of a 10 V programmable Josephson voltage standard (PJVS) array operated in a cryocooler. The operating margins at 70 GHz frequencies are discussed and compared to liquid helium cooling. A crucial factor is the limited heat conduction between PJVS chip and cold stage. A method is presented to estimate the heat conduction directly with the Josephson array. Finally, a direct comparison between two 10 V PJVS arrays, one operated in a cryocooler and the other cooled with liquid helium is performed and shows an agreement to a level of 1 part in 10^\{−10\}. |
Zitierung
Starkloff, M., Schubert, M., Lee, J., Behr, R., Palafox Gamir, L., Böck, A. C., Fleischmann, P. M., & Peiselt, K. (2016). Dry cooling of a 10 V programmable Josephson voltage standard array. Conference on Precision Electromagnetic Measurements (CPEM) 2016, Ottawa, 10-15, July, 2016, Canada. https://doi.org/10.1109/cpem.2016.7540472