Titel New X-ray parallel beam facility XPBF 2.0 for the characterization of silicon pore optics
Autor(in); Institution
Krumrey, Michael; 7.1, Radiometrie mit Synchrotronstrahlung, PTB-Berlin
Müller, Peter; 7.1, Radiometrie mit Synchrotronstrahlung, PTB-Berlin
Cibik, Levent; 7.1, Radiometrie mit Synchrotronstrahlung, PTB-Berlin
Collon, Max; cosine Research B.V., Warmond, THE NETHERLANDS
Vacanti, Giuseppe; cosine Research B.V., Warmond, THE NETHERLANDS
Bavdaz, Marcos; European Space Agency, ESTEC, Noordwijk, THE NETHERLANDS
Quelle/Jahr Space Telescopes and Instrumentation 2016: Ultraviolet to Gamma Ray:(2016), 9905N-1 - 9905N-8
Schriftenreihe Proceedings of SPIE: 9905
Herausgeber(in)
den Herder, Jan-Willem A.
Takahashi, T.
Bautz, Marshall
ISSN 0277-786X
ISBN 978-1-5106-0189-5
DOI
Verlag Bellingham, Wash.: SPIE
Konferenzangaben SPIE Astronomical Telescopes + Instrumentation, Edinburgh, 26 June - 1 July 2016, UK
Freie Schlagworte ATHENA ; Silicon Pore Optics (SPO) ; synchrotron radiation ; XPBF
Zusammenfassung A new X-ray parallel beam facility (XPBF 2.0) has been installed in the laboratory of the Physikalisch-Technische Bundesanstalt at the synchrotron radiation facility BESSY II in Berlin to characterize silicon pore optics (SPOs) for the future X-ray observatory ATHENA. As the existing XPBF which is operated since 2005, the new beamline provides a pencil beam of very low divergence, a vacuum chamber with a hexapod system for accurate positioning of the SPO to be investigated, and a vertically movable CCD-based camera system to register the direct and the reflected beam. In contrast to the existing beamline, a multilayer-coated toroidal mirror is used for beam monochromatization at 1.6 keV and collimation, enabling the use of beam sizes between about 100 µm and at least 5 mm. Thus the quality of individual pores as well as the focusing properties of large groups of pores can be investigated. The new beamline also features increased travel ranges for the hexapod to cope with larger SPOs and a sample to detector distance of 12 m corresponding to the envisaged focal length of ATHENA.

Zitierung

Krumrey, M., Müller, P., Cibik, L., Collon, M., Vacanti, G., & Bavdaz, M. (2016). New X-ray parallel beam facility XPBF 2.0 for the characterization of silicon pore optics. SPIE Astronomical Telescopes + Instrumentation, Edinburgh, 26 June - 1 July 2016, UK. https://doi.org/10.1117/12.2231687