Zugriffsnummer 36136
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel Detailing local multi-channel RF surface coil versus body RF coil transmission for cardiac MRI at 3 Tesla: which configuration is winning the game?
Autor(in); Institution
Weinberger, O.; Berlin Ultrahigh Field Facility (BUFF), Max Delbrueck Center for Molecular Medicine (MDC), Berlin, GERMANY
Winter, L.; Berlin Ultrahigh Field Facility (BUFF), Max Delbrueck Center for Molecular Medicine (MDC), Berlin, GERMANY
Dieringer, M.A.; Berlin Ultrahigh Field Facility (BUFF), Max Delbrueck Center for Molecular Medicine (MDC), Berlin, GERMANY
Els, A.; Berlin Ultrahigh Field Facility (BUFF), Max Delbrueck Center for Molecular Medicine (MDC), Berlin, GERMANY
Oezerdem, C.; Berlin Ultrahigh Field Facility (BUFF), Max Delbrueck Center for Molecular Medicine (MDC), Berlin, GERMANY
Cassara, Antonino; 8.1, Medizinische Messtechnik, PTB-Berlin
Pfeiffer, Harald; 8.1, Medizinische Messtechnik, PTB-Berlin
Niendorf, Th.; Berlin Ultrahigh Field Facility (B.U.F.F.), Max Delbrück Center for Molecular Medicine in the Helmholtz Association, Berlin, GERMANY
Quelle/Jahr ISMRM 24th annual meeting & exhibition: 07 - 13 May 2016, Singapore:(2016), 4 S.
Availability [USB-Stick]
Verlag Berkeley, Calif.: ISMRM
Konferenzangaben 24th ISMRM Annual Meeting & Exhibition, Singapore, 07 May – 13 May , 2016, Singapore
Zusammenfassung In this work a local four-channel transmit/receive RF coil dedicated for cardiac MR at 3T is compared to a conventional built-in body RF coil in conjunction with a four-channel receive-only RF coil. SAR and B1+ simulations of both configurations are shown. The invivo efficiency performance of both coils in respect to B1+/sqrt(SAR) is demonstrated in 12 healthy subjects. The efficiency surplus of the local RF coil was used to increase the applicable flip angle FASSFP of a standard high resolution 2D SSFP protocol or to shorten the used repetition time TRSSFP by 54%.

Zitierung

Weinberger, O., Winter, L., Dieringer, M., Els, A., Oezerdem, C., Cassara, A., Pfeiffer, H., & Niendorf, T. (2016). Detailing local multi-channel RF surface coil versus body RF coil transmission for cardiac MRI at 3 Tesla: which configuration is winning the game? 24th ISMRM Annual Meeting & Exhibition, Singapore, 07 May – 13 May , 2016, Singapore.